Pineda Federico D, Medved Milica, Fan Xiaobing, Karczmar Gregory S
Department of Radiology, University of Chicago, Chicago, Illinois, USA.
Magn Reson Med. 2016 Apr;75(4):1565-73. doi: 10.1002/mrm.25751. Epub 2015 May 27.
To develop a method for mapping the B1 field using a reference signal from a tissue with known T1.
Flip angle correction factors were calculated in a region with a known "gold standard" T1; by comparing T1 values from a variable flip angle (VFA) sequence to the "gold standard" and correcting the value of the Ernst angle. The resulting partial B1 map was interpolated for all other regions. In the breast, fat is an ideal reference tissue because its T1 is spatially homogeneous and interpatient variability is low. This method was tested with scans of phantoms and patients (n = 4) on a 3T magnet. The performance of the method was evaluated by comparing the results of VFA T1 mapping with and without B1 correction to inversion recovery (IR) T1 maps.
Phantom data determined that a linear inverse distance weighted interpolation accurately recovered the full B1 map. Use of interpolated maps to correct the VFA data in vivo, reduced the average difference in the T1 of parenchyma between VFA and IR results from 58% to 8%.
This proof-of-principle study showed that it is possible to recover a full and accurate map of the B1 field in the breast by using a reference tissue (fat) with an accurately measured T1.
开发一种利用来自具有已知T1值的组织的参考信号来绘制B1场的方法。
在具有已知“金标准”T1的区域中计算翻转角校正因子;通过将可变翻转角(VFA)序列的T1值与“金标准”进行比较并校正恩斯特角的值。对所有其他区域对所得的部分B1图进行插值。在乳腺中,脂肪是理想的参考组织,因为其T1在空间上是均匀的且患者间变异性低。该方法在3T磁体上对体模和患者(n = 4)进行扫描测试。通过比较有和没有B1校正的VFA T1映射结果与反转恢复(IR)T1映射来评估该方法的性能。
体模数据确定线性逆距离加权插值能准确恢复完整的B1图。在体内使用插值图校正VFA数据,将VFA和IR结果之间实质T1的平均差异从58%降低到8%。
这项原理验证研究表明,通过使用具有精确测量T1的参考组织(脂肪),有可能恢复乳腺中完整且准确的B1场图。