Abstract

An investigation to study changes in the metal surfaces and the polyurethane insulation of heart pacemaker leads under controlled in vitro conditions was conducted. A polyurethane (Pellethane 2363-80A)/Co-Ni-Cr-Mo (MP35N) wire lead was exposed in Hanks' physiological saline solution for 14 months and then analyzed using scanning electron microscopy, x-ray energy dispersive analysis, and small angle x-ray scattering. Results showed that some leakage of solution into the lead had occurred and changes were present on both the metal and the polyurethane surfaces.

摘要

开展了一项研究,以在可控的体外条件下,研究心脏起搏器导线的金属表面和聚氨酯绝缘层的变化。将一根聚氨酯(聚醚氨酯2363-80A)/钴镍铬钼(MP35N)金属丝导线置于汉克斯生理盐溶液中14个月,然后使用扫描电子显微镜、X射线能谱分析和小角X射线散射进行分析。结果表明,溶液已发生一定程度的泄漏进入导线,金属和聚氨酯表面均出现了变化。