Suppr超能文献

双翻转角 IR-FLASH 结合自旋历史映射用于 B1+校正的 T1 映射:在 T1 心血管磁共振多任务中的应用。

Dual flip-angle IR-FLASH with spin history mapping for B1+ corrected T1 mapping: Application to T1 cardiovascular magnetic resonance multitasking.

机构信息

Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.

Siemens Medical Solutions USA, Inc., Los Angeles, California, USA.

出版信息

Magn Reson Med. 2021 Dec;86(6):3182-3191. doi: 10.1002/mrm.28935. Epub 2021 Jul 26.

Abstract

PURPOSE

To develop a single-scan method for -corrected T mapping and apply it for free-breathing (FB) cardiac MR multitasking without electrocardiogram (ECG) triggering.

METHODS

One dual flip-angle (2FA) inversion recovery (IR)-FLASH scan provides two observations of (apparent T ) corresponding to two distinct combinations of the nominal FA α and . Spatiotemporally coregistered T and spin history maps are obtained by fitting the 2FA signal model. T estimate accuracy and repeatability for single flip-angle (1FA) and 2FA IR-FLASH sequence MR multitasking were evaluated at 3T. A T phantom was first imaged on the scanner table, then on two human subjects' thoraxes in both breath-hold (BH) and FB conditions. IR-turbo spin echo (IR-TSE) static phantom T measurements served as reference. In 10 healthy subjects, myocardial T was evaluated with ECG-free, FB multitasking sequences alongside ECG-triggered BH MOLLI.

RESULTS

For phantom-on-table T estimates, 2FA agreed better with IR-TSE (intraclass correlation coefficient [ICC] = 0.996, mean error ± SD = -1.6% ± 1.9%) than did 1FA (ICC = 0.922; mean error ± SD = -4.3% ± 12%). For phantom-on-thorax, 2FA was more repeatable and robust to respiration than 1FA (coefficient of variation [CoV] = 1.2% 2FA, = 11.3% 1FA). In vivo, in intrasession T repeatability, 2FA (septal CoV = 2.4%, six-segment CoV = 4.4%) outperformed 1FA (septal CoV = 3.1%, six-segment CoV = 5.5%). In six-segment T homogeneity, 2FA (CoV = 7.9%) also outperformed 1FA (CoV = 11.1%).

CONCLUSION

The 2FA IR-FLASH improves T estimate accuracy and repeatability over 1FA IR-FLASH, and enables single-scan -corrected T mapping without BHs or ECG when used with MR multitasking.

摘要

目的

开发一种用于校正 T 映射的单次扫描方法,并将其应用于无需心电图 (ECG) 触发的自由呼吸 (FB) 心脏磁共振多任务成像。

方法

单次双翻转角 (2FA) 反转恢复 (IR)-FLASH 扫描可提供两个对应于名义翻转角 (FA)α和的两个不同组合的观测值(表观 T)。通过拟合 2FA 信号模型,获得时空配准的 T 和自旋历史图。在 3T 上评估了单翻转角 (1FA) 和 2FA IR-FLASH 序列多任务成像的 T 估计准确性和可重复性。首先在扫描仪台上对 T 体模进行成像,然后在两个人体胸部进行屏气 (BH) 和 FB 两种条件下的成像。IR-涡轮自旋回波 (IR-TSE) 静态体模 T 测量作为参考。在 10 名健康受试者中,使用无 ECG、FB 多任务成像序列评估心肌 T,同时进行 ECG 触发 BH MOLLI。

结果

对于体模台上的 T 估计值,2FA 与 IR-TSE 的一致性更好(组内相关系数 [ICC] = 0.996,平均误差 ± SD = -1.6% ± 1.9%),而 1FA 的一致性较差(ICC = 0.922;平均误差 ± SD = -4.3% ± 12%)。对于体模在胸部的情况,2FA 比 1FA 更具可重复性且对呼吸更稳健(变异性 [CoV] = 1.2% 2FA,= 11.3% 1FA)。在体内,在单次扫描 T 重复性方面,2FA(间隔 CoV = 2.4%,六节段 CoV = 4.4%)优于 1FA(间隔 CoV = 3.1%,六节段 CoV = 5.5%)。在六节段 T 均匀性方面,2FA(CoV = 7.9%)也优于 1FA(CoV = 11.1%)。

结论

2FA IR-FLASH 提高了 T 估计的准确性和重复性,优于 1FA IR-FLASH,并且当与磁共振多任务成像结合使用时,可以在不进行 BH 和 ECG 的情况下实现单次扫描校正 T 映射。

相似文献

引用本文的文献

10
Cardiac MR: From Theory to Practice.心脏磁共振成像:从理论到实践。
Front Cardiovasc Med. 2022 Mar 3;9:826283. doi: 10.3389/fcvm.2022.826283. eCollection 2022.

本文引用的文献

3
Fast myocardial T mapping using cardiac motion correction.使用心脏运动校正的快速心肌 T 映射。
Magn Reson Med. 2020 Feb;83(2):438-451. doi: 10.1002/mrm.27935. Epub 2019 Aug 16.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验