To assess (1) step count accuracy of the Fitbit Zip, compared to manual step count, in people receiving outpatient rehabilitation, in indoor and outdoor conditions, and (2) impact of slow walking speed on Fitbit accuracy.
Observational study.
A metropolitan rehabilitation hospital.
Adults ( = 88) attending a subacute rehabilitation outpatient clinic with walking speeds of between 0.4 and 1.0 m/s.
Two 2-minute walk tests, one indoors and one outdoors, completed in random order.
Step count recorded manually by observation and by a Fitbit Zip, attached to the shoe on the dominant or non-affected side. Subgroup analysis included assessment accuracy for those considered limited community walkers (slower than 0.8 m/s) and those considered community walkers (faster than 0.8 m/s).
The Fitbit significantly ( < 0.05) undercounted steps compared to manual step count, indoors and outdoors, with percentage agreement slightly higher outdoors (mean 92.4%) than indoors (90.1%). Overall, there was excellent consistent agreement between the Fitbit and manual step count for both indoor (ICC 0.83) and outdoor (ICC 0.88) walks. The accuracy of the Fitbit was significantly ( < 0.05) reduced in those who walked slower than 0.8 m/s outdoors (ICC 0.80) compared to those who walk faster than 0.8 m/s (ICC 0.90).
The Fitbit Zip shows high step count accuracy with manual step count in a mixed subacute rehabilitation population. However, accuracy is affected by walking speed, with decreased accuracy in limited community walkers.
评估(1)在接受门诊康复治疗的人群中,Fitbit Zip 计步器的计步准确性,分别在室内和室外环境下,并(2)慢步行速度对 Fitbit 准确性的影响。
观察性研究。
一家都会区康复医院。
步速在 0.4 至 1.0 m/s 之间、参加亚急性康复门诊的成年人(=88 人)。
以随机顺序各完成 2 分钟的步行测试,一次在室内,一次在室外。
手动观察和通过附着在优势或非患侧鞋上的 Fitbit Zip 记录的步数。亚组分析包括对被认为是有限社区步行者(慢于 0.8 m/s)和社区步行者(快于 0.8 m/s)的评估准确性。
与手动计步相比,Fitbit 在室内和室外均显著(<0.05)低估了步数,室外的百分比一致性略高于室内(平均 92.4%对 90.1%)。总体而言,Fitbit 与手动计步在室内(ICC 0.83)和室外(ICC 0.88)步行中具有极好的一致性。在室外步速慢于 0.8 m/s 的人群中(ICC 0.80),Fitbit 的准确性显著(<0.05)降低,而在步速快于 0.8 m/s 的人群中(ICC 0.90)则没有。
在混合亚急性康复人群中,Fitbit Zip 与手动计步相比显示出较高的计步准确性。然而,准确性受到步行速度的影响,在有限的社区步行者中准确性降低。