Accuracy of the FitBit at walking speeds and cadences relevant to clinical rehabilitation populations.

作者信息

Singh Abhishek K, Farmer Craig, Van Den Berg Maayken L E, Killington Maggie, Barr Christopher J

机构信息

Department of Rehabilitation Aged and Extended Care, School of Health Sciences, Flinders University, South Australia, Australia.

Department of Physiotherapy, Repatriation General Hospital, South Australia, Australia.

出版信息

Disabil Health J. 2016 Apr;9(2):320-3. doi: 10.1016/j.dhjo.2015.10.011. Epub 2015 Dec 1.

Abstract

BACKGROUND

Activity monitoring is used for motivation and assessing patients in the rehabilitation setting. Monitors available on the market have not been assessed for accuracy at walking patterns seen in the clinical setting.

OBJECTIVES

To assess the accuracy of the Fitbit Zip device at different anatomical locations at walking speeds, step length, and cadence similar to those seen in clinical rehabilitation populations.

METHODS

Ten healthy participants completed 3 sets of 12 combinations of speed and cadence ranging from 0.2 m/s at 48 steps per minute to 1.0 m/s at 120 steps per minute along a 20 m walkway. Step length was controlled by ground markings and cadence was controlled by a metronome. Five Fitbit Zip devices were concurrently worn at the chest, hip, shin, ankle and forefoot. Percent accuracy for each location and walking condition combination were calculated.

RESULTS

At the chest and hip the Fitbit has poor accuracy below 0.7 m/s, however when worn distally, can be accurate to 5.5% at speeds as slow as 0.5 m/s.

CONCLUSIONS

For patient populations with slow walking speed, activity monitoring with the Fitbit can be achieved provided the device is located distally on the lower limb.

摘要

背景

活动监测用于康复环境中的激励和患者评估。市场上现有的监测器尚未针对临床环境中观察到的步行模式的准确性进行评估。

目的

评估Fitbit Zip设备在不同解剖位置对于类似于临床康复人群中观察到的步行速度、步长和步频的准确性。

方法

10名健康参与者沿着一条20米的走道完成了3组,每组12种速度和步频组合,速度范围从每分钟48步时的0.2米/秒到每分钟120步时的1.0米/秒。步长由地面标记控制,步频由节拍器控制。同时在胸部、臀部、小腿、脚踝和前脚掌佩戴5个Fitbit Zip设备。计算每个位置和步行条件组合的准确率。

结果

在胸部和臀部,Fitbit在低于0.7米/秒时准确性较差,然而,当佩戴在肢体远端时,在低至0.5米/秒的速度下准确率可达5.5%。

结论

对于步行速度较慢的患者群体,只要将Fitbit设备佩戴在下肢远端,就可以实现活动监测。