[陶瓷髋关节假体的植入物骨折]

[Implant fractures in ceramic hip endoprostheses].

作者信息

Heisel J, Schmitt E

机构信息

Oberarzt der Orthopädischen Universitätsklinik, Homburg/Saar.

出版信息

Z Orthop Ihre Grenzgeb. 1987 Sep-Oct;125(5):480-90. doi: 10.1055/s-2008-1044744.

Abstract

Between 1974 and 1987, a total of 2356 ceramic hip endoprostheses, both primary and secondary, were implanted at the Orthopedic University Clinic in Homburg/Saar. Strem fractures occurred in three of these patients only, which corresponds to a relative frequency of 0.13%. In two instances, these were fatigue fractures of the AUTOPHOR I (WISIL) prosthesis stem, while in the third instance the fracture, of stem type 2 (ENDOCAST), was the result of a fault during casting. Fractures of ceramic components as a result of a direct trauma of sufficient force were observed in six cases (relative frequency of 0.25%). Persistent fractures only occurred where there was recurrent knocking of a long ceramic collar against the rim of the acetabulum (eight cases, relative frequency of 0.34%), mainly as a result of primary or secondary false positioning of the acetabulum (surgical error, aseptic loosening). Ceramic fractures when using the ceramics/polyethylene combination for joints have yet to be recorded. Given the authors' 12 years of experience, and considering the large number of patients treated fractures of high quality BIOLOX ceramics are a rarity. Both persistent fractures and exaggerated ceramic abrasion were almost always the result of false positioning of the acetabulum, with the recurrent knocking of the hip head ceramic collar playing a significant role. It is evident that the extent of tolerance of ceramics as regards undue strain with a subsequent danger of fracture is lower than that of the metal-polyethylene combination.

摘要

1974年至1987年间,在洪堡/萨尔的骨科大学诊所共植入了2356个陶瓷髋关节假体,包括初次植入和二次植入的。仅3例患者发生了柄部骨折,相对发生率为0.13%。其中2例是AUTOPHOR I(WISIL)假体柄的疲劳骨折,而第3例2型柄(ENDOCAST)骨折是铸造过程中的缺陷所致。6例(相对发生率0.25%)观察到由于足够力量的直接创伤导致陶瓷部件骨折。持续性骨折仅发生在长陶瓷颈反复撞击髋臼边缘的情况下(8例,相对发生率0.34%),主要是由于髋臼的初次或二次错误定位(手术失误、无菌性松动)。使用陶瓷/聚乙烯关节组合时尚未记录到陶瓷骨折。根据作者12年的经验,并考虑到大量接受治疗的患者,高质量BIOLOX陶瓷骨折实属罕见。持续性骨折和过度的陶瓷磨损几乎总是髋臼错误定位的结果,股骨头陶瓷颈的反复撞击起了重要作用。显然,陶瓷在承受过度应变及随后骨折风险方面的耐受程度低于金属-聚乙烯组合。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验