Grønbæk-Thorsen Freja, Jensen Camilla, Østergaard Jesper, Møller Laura Hyrup, Gammelgaard Bente
Department of Pharmacy, University of Copenhagen, Universitetsparken 2, 2100, Copenhagen, Denmark.
Anal Bioanal Chem. 2021 Nov;413(26):6479-6488. doi: 10.1007/s00216-021-03611-1. Epub 2021 Aug 30.
In the present study, a method for quantitation of the pharmaceutical peptide oxytocin (OT) and its diselenide-containing analogue (SeOT) in human plasma was developed using gradient elution LC-ICP-MS/MS. Plasma samples were precipitated with acetonitrile containing 1.0% TFA in a volume ratio of 1+3 (sample+precipitation agent) before analysis. Post-column isotope dilution analysis (IDA) was applied for quantitation and was compared with external calibration. Both calibration methods appeared to be fit for purpose regarding figures of merit including linearity, precision, LOD, LOQ and recovery. Analysis of OT and SeOT showed that selenium-based analysis is considerably more sensitive and selective compared to the sulfur-based analysis. Despite the relatively simpler setup of external calibration, IDA can be advantageous because it compensates for instrument drift and changes in organic solvent concentration. The method was applied for a stability study showing the degradation of OT and SeOT in plasma. The degradation of SeOT was faster than the degradation of OT in plasma. Thus, possible stability effects should be considered before replacing a disulfide bridge with a diselenide bridge or introducing a diselenide label in a potential drug.
在本研究中,采用梯度洗脱液相色谱 - 电感耦合等离子体质谱联用技术(LC - ICP - MS/MS)建立了一种定量测定人血浆中药物肽催产素(OT)及其含二硒化物类似物(SeOT)的方法。分析前,血浆样品用含1.0%三氟乙酸(TFA)的乙腈按1 + 3的体积比(样品 + 沉淀剂)进行沉淀。柱后同位素稀释分析(IDA)用于定量,并与外标法进行比较。就线性、精密度、检测限(LOD)、定量限(LOQ)和回收率等分析性能指标而言,两种校准方法似乎都适用。对OT和SeOT的分析表明,与基于硫的分析相比,基于硒的分析灵敏度和选择性要高得多。尽管外标法的设置相对简单,但IDA可能更具优势,因为它可以补偿仪器漂移和有机溶剂浓度的变化。该方法用于稳定性研究,结果表明OT和SeOT在血浆中会发生降解。SeOT在血浆中的降解速度比OT快。因此,在用二硒化物桥取代二硫键或在潜在药物中引入二硒化物标签之前,应考虑可能的稳定性影响。