Suppr超能文献

用于评估血浆催产素的高灵敏度定量液相色谱-串联质谱检测方法的开发与验证。

Development and validation of a highly-sensitive, quantitative LC-MS/MS assay to evaluate plasma oxytocin.

作者信息

Grifnée E, Mackowiak A, Demeuse J, Schoumacher M, Huyghebaert L, Determe W, Dubrowski T, Massonnet P, Peeters S, Scantamburlo G, Cavalier E, Goff C Le

机构信息

Department of Clinical Chemistry, University Hospital of Liège, Belgium.

Department of Clinical Chemistry, CIRM, University of Liège, Belgium.

出版信息

J Mass Spectrom Adv Clin Lab. 2025 Feb 22;36:19-28. doi: 10.1016/j.jmsacl.2025.02.002. eCollection 2025 Apr.

Abstract

INTRODUCTION

Oxytocin is a 9-amino acid peptide that serves as neuromodulator in the human central nervous system. This peptide is implicated in the regulation of diverse behaviors and plays a significant role in positive social interaction. Currently, oxytocin levels are measured using immunoassays. However, these methods have several limitations that can lead to false results and erroneous interpretation. Given the remarkably low endogenous level of oxytocin in human plasma (low ng/L levels), we developed and rigorously validated a novel and highly sensitive LC-MS/MS method for oxytocin quantification in plasma.

METHODS

Oxytocin was initially extracted using solid-phase extraction with an Oasis HLB 30 mg plate and then subjected to LC-MS/MS analysis. PBS-0.1 % BSA served as surrogate matrix for the preparation of validation samples and the calibration curve, ensuring no endogenous interference. The validation design followed the Clinical Laboratory Standards Institute guidelines. Precision, accuracy, and measurement uncertainty were determined using single-nested analysis of variance and e.noval software.

RESULTS

A lower limit of quantification of 1 ng/L was achieved. The method was validated for oxytocin concentrations ranging from 1 ng/L to 75 ng/L, with precision (coefficient of variation) below 10 %, accuracy ranging from 94 % to 108 %, and measurement uncertainty below 15 %.

CONCLUSION

In this work, we developed and validated a highly sensitive LC-MS/MS method for the quantification of oxytocin in plasma. Our novel methodology is well-suited for clinical applications.

摘要

引言

催产素是一种由9个氨基酸组成的肽,在人类中枢神经系统中作为神经调节剂发挥作用。这种肽参与多种行为的调节,在积极的社会互动中发挥重要作用。目前,催产素水平通过免疫测定法进行测量。然而,这些方法存在一些局限性,可能导致错误结果和错误解释。鉴于人血浆中催产素的内源性水平极低(低至纳克/升水平),我们开发并严格验证了一种用于血浆中催产素定量的新型高灵敏度液相色谱-串联质谱(LC-MS/MS)方法。

方法

催产素首先使用Oasis HLB 30毫克板进行固相萃取,然后进行LC-MS/MS分析。PBS-0.1%牛血清白蛋白用作制备验证样品和校准曲线的替代基质,确保无内源性干扰。验证设计遵循临床实验室标准协会的指南。使用单因素方差分析和e.noval软件确定精密度、准确度和测量不确定度。

结果

实现了1纳克/升的定量下限。该方法在1纳克/升至75纳克/升的催产素浓度范围内得到验证,精密度(变异系数)低于10%,准确度在94%至108%之间,测量不确定度低于15%。

结论

在这项工作中,我们开发并验证了一种用于血浆中催产素定量的高灵敏度LC-MS/MS方法。我们的新方法非常适合临床应用。