Department of Radiation Oncology (Maastro), GROW School for Oncology, Maastricht University Medical Centre+, Maastricht, The Netherlands.
Department of Radiation Oncology (Maastro), GROW School for Oncology, Maastricht University Medical Centre+, Maastricht, The Netherlands.
Radiother Oncol. 2021 Oct;163:136-142. doi: 10.1016/j.radonc.2021.08.014. Epub 2021 Aug 27.
Quality of automatic contouring is generally assessed by comparison with manual delineations, but the effect of contour differences on the resulting dose distribution remains unknown. This study evaluated dosimetric differences between treatment plans optimized using various organ-at-risk (OAR) contouring methods.
OARs of twenty lung cancer patients were manually and automatically contoured, after which user-adjustments were made. For each contour set, an automated treatment plan was generated. The dosimetric effect of intra-observer contour variation and the influence of contour variations on treatment plan evaluation and generation were studied using dose-volume histogram (DVH)-parameters for thoracic OARs.
Dosimetric effect of intra-observer contour variability was highest for Heart D (3.4 ± 6.8 Gy) and lowest for Lungs-GTV D (0.3 ± 0.4 Gy). The effect of contour variation on treatment plan evaluation was highest for Heart D (6.0 ± 13.4 Gy) and Esophagus D (8.7 ± 17.2 Gy). Dose differences for the various treatment plans, evaluated on the reference (manual) contour, were on average below 1 Gy/1%. For Heart D, higher dose differences were found for overlap with PTV (median 0.2 Gy, 95% 1.7 Gy) vs. no PTV overlap (median 0 Gy, 95% 0.5 Gy). For D-parameters, largest dose difference was found between 0-1 cm distance to PTV (median 1.5 Gy, 95% 4.7 Gy).
Dose differences arising from automatic contour variations were of the same magnitude or lower than intra-observer contour variability. For Heart D, we recommend delineation errors to be corrected when the heart overlaps with the PTV. For D-parameters, we recommend checking contours if the distance is close to PTV (<5 cm). For the lungs, only obvious large errors need to be adjusted.
自动勾画的质量通常通过与手动勾画的比较来评估,但勾画差异对最终剂量分布的影响尚不清楚。本研究评估了使用不同危及器官(OAR)勾画方法优化的治疗计划之间的剂量学差异。
对 20 例肺癌患者的 OAR 进行手动和自动勾画,然后进行用户调整。为每个轮廓集生成一个自动化的治疗计划。使用胸部 OAR 的剂量-体积直方图(DVH)参数研究了观察者内轮廓变化的剂量学效应以及轮廓变化对治疗计划评估和生成的影响。
观察者内轮廓变异性的剂量学效应在心脏 D(3.4±6.8 Gy)最高,在肺部-GTV D(0.3±0.4 Gy)最低。轮廓变化对治疗计划评估的影响在心脏 D(6.0±13.4 Gy)和食管 D(8.7±17.2 Gy)最高。在参考(手动)轮廓上评估的各种治疗计划之间的剂量差异平均低于 1 Gy/1%。对于心脏 D,与无 PTV 重叠(中位数 0 Gy,95% 0.5 Gy)相比,与 PTV 重叠(中位数 0.2 Gy,95% 1.7 Gy)的剂量差异更高。对于 D 参数,与 PTV 距离为 0-1 cm 时发现最大的剂量差异(中位数 1.5 Gy,95% 4.7 Gy)。
自动轮廓变化引起的剂量差异与观察者内轮廓变化相同或更小。对于心脏 D,建议当心脏与 PTV 重叠时纠正心脏勾画误差。对于 D 参数,如果距离接近 PTV(<5 cm),建议检查轮廓。对于肺部,只有明显的大误差需要调整。