Suppr超能文献

Micra-AV 无导线起搏器与房室(传导)失同步:一个逐步的过程。

Micra-AV leadless pacemaker and atrioventricular (dys)synchrony: A stepwise process.

机构信息

Cardiology Unit, ASST-Fatebenefratelli Sacco, Luigi Sacco University Hospital, Milan, Italy.

出版信息

Pacing Clin Electrophysiol. 2021 Oct;44(10):1738-1742. doi: 10.1111/pace.14352. Epub 2021 Sep 12.

Abstract

Leadless pacemakers have become a major breakthrough in the management of bradyarrhythmia as an attractive alternative to the standard transvenous pacemakers, but safety and long-term outcomes of the new Micra-AV, which is capable guarantee atrioventricular (AV) synchrony, have been poorly described. Hence, we describe how we managed a 57-year-old patient who developed pacemaker syndrome due to significant Micra-AV dyssynchrony. This case emphasizes how leadless pacemaker AV synchrony could be overestimated, thereby requiring a stepwise process leading to adequate device reprogramming. Holter-ECG monitoring and exercise test resulted to be valuable tools for an early detection of inadequate AV synchrony, integrating and completing device reports.

摘要

无导线起搏器已成为治疗缓律失常的重大突破,作为传统经静脉起搏器的一种有吸引力的替代方案,但新的 Micra-AV (能够保证房室(AV)同步)的安全性和长期结果描述甚少。因此,我们描述了如何处理一位 57 岁患者,该患者因明显的 Micra-AV 不同步而出现起搏器综合征。该病例强调了无导线起搏器 AV 同步性可能被高估,从而需要逐步进行适当的设备重新编程。动态心电图监测和运动试验是早期发现 AV 同步不足的有价值的工具,可整合并完善设备报告。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验