Ayan Ahmet S, Kim Grace, Whitaker Matthew, Al-Hallaq Hania, Hsu Shu-Hui, Woollard Jeffrey, Roberts Donald A, Shtraus Natan, Gao Song, Gupta Nilendu, Moran Jean M
Ohio State University, OH, United States of America.
UC-San Diego, CA, United States of America.
Biomed Phys Eng Express. 2021 Oct 5;7(6). doi: 10.1088/2057-1976/ac2876.
The purpose of this study was to develop and evaluate a framework to support automated standardized testing and analysis of Cone Beam Computed Tomography (CBCT) image quality QA across multiple institutions. A survey was conducted among the participating institutions to understand the variability of the CBCT QA practices. A commercial, automated software platform was validated by seven institutions participating in a consortium dedicated to automated quality assurance. The CBCT image analysis framework was used to compare periodic QA results among 23 linear accelerators (linacs) from seven institutions. The CBCT image quality metrics (geometric distortion, spatial resolution, contrast, HU constancy, uniformity and noise) data are plotted as a function of means with the upper and lower control limits compared to the linac acceptance criteria and AAPM recommendations. For example, mean geometric distortion and HU constancy metrics were found to be 0.13 mm (TG142 recommendation: ≤2 mm) and 13.4 respectively (manufacturer acceptance specification: ≤±50).Image upload and analysis process was fully automated using a MATLAB-based platform. This analysis enabled a quantitative, longitudinal assessment of the performance of quality metrics which were also compared across 23 linacs. For key CBCT parameters such as uniformity, contrast, and HU constancy, all seven institutions used stricter goals than what would be recommended based on the analysis of the upper and lower control limits. These institutional goals were also found to be stricter than that found in AAPM published guidance. This work provides a reference that could be used to machine-specific optimized tolerance of CBCT image maintenance via control charts to monitor performance we well as the sensitivity of different tests in support of a broader quality assurance program. To ensure the daily image quality needed for patient care, the optimized statistical QA metrics recommended to using along with risk-based QA.
本研究的目的是开发并评估一个框架,以支持多个机构对锥束计算机断层扫描(CBCT)图像质量QA进行自动化标准化测试和分析。对参与机构进行了一项调查,以了解CBCT QA实践的变异性。一个商业自动化软件平台由参与一个致力于自动化质量保证的联盟的七个机构进行了验证。CBCT图像分析框架用于比较来自七个机构的23台直线加速器(直线加速器)的定期QA结果。CBCT图像质量指标(几何畸变、空间分辨率、对比度、HU一致性、均匀性和噪声)数据作为均值的函数绘制,并将上下控制限与直线加速器验收标准和AAPM建议进行比较。例如,发现平均几何畸变和HU一致性指标分别为0.13毫米(TG142建议:≤2毫米)和13.4(制造商验收规范:≤±50)。图像上传和分析过程使用基于MATLAB的平台完全自动化。该分析能够对质量指标的性能进行定量纵向评估,这些指标也在23台直线加速器之间进行了比较。对于关键的CBCT参数,如均匀性、对比度和HU一致性,所有七个机构使用的目标都比根据上下控制限分析建议的目标更严格。这些机构目标也比AAPM发布的指南中发现的目标更严格。这项工作提供了一个参考,可用于通过控制图对CBCT图像维护进行特定于机器的优化公差,以监测性能以及不同测试的灵敏度,以支持更广泛的质量保证计划。为确保患者护理所需的每日图像质量,建议使用优化的统计QA指标以及基于风险的QA。