通过锥体像差测量法研究多焦点、可调节眼内晶状体和扩展景深人工晶状体的视网膜图像质量。

Retinal image quality with multifocal, EDoF, and accommodative intraocular lenses as studied by pyramidal aberrometry.

作者信息

Alio Jorge L, D'Oria Francesco, Toto Francesca, Balgos Joan, Palazon Antonio, Versaci Francesco, Alio Del Barrio Jorge L

机构信息

Vissum Innovation, Alicante, Spain.

Division of Ophthalmology, Universidad Miguel Hernández, Vissum Miranza, c/ Cabañal, 1, 03016, Alicante, Spain.

出版信息

Eye Vis (Lond). 2021 Oct 6;8(1):37. doi: 10.1186/s40662-021-00258-y.

Abstract

BACKGROUND

To study and compare the clinical optical image quality following implantation with different premium IOLs by analysing the point spread function (PSF) Strehl ratio using a pyramidal wavefront sensor (PWS)-based aberrometer.

METHODS

This study included 194 eyes implanted with: (a) 19 AcrySof SA60AT (control group); (b) 19 Miniwell; (c) 24 LENTIS Mplus LS-313 MF30; d) 33 LENTIS Mplus LS-313 MF15; (e) 17 AkkoLens Lumina; (f) 31 AT LISA Tri 839MP; (g) 20 Precizon Presbyopic; (h) 20 AcrySof IQ PanOptix; (i) 11 Tecnis Eyhance. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root mean square (RMS), low-order aberration (LOA) and high-order aberration (HOA) RMS measured by PWS aberrometer.

RESULTS

AT LISA Tri had the highest PSFw2 Strehl ratio at both 3.0- and 4.0-mm pupil size (0.52 ± 0.14 and 0.31 ± 0.10; P < 0.05), followed by SA60AT (0.41 ± 0.11 and 0.28 ± 0.07) and PanOptix (0.4 ± 0.07 and 0.26 ± 0.04). AT LISA Tri was found to provide a significantly better retinal image quality than PanOptix at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.004). Mplus MF15 was found to be significantly better than Mplus MF30 at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.002). Total RMS, LOA RMS, HOA RMS, PSF Strehl ratio and PSFw2 varied significantly between the studied groups (P < 0.001).

CONCLUSIONS

Far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens. AT LISA Tri, SA60AT and PanOptix showed the highest values of far distance retinal image quality, while the lowest PSFw2 Strehl ratios were displayed by Miniwell, Mplus MF30 and Precizon Presbyopic.

摘要

背景

通过使用基于金字塔波前传感器(PWS)的像差仪分析点扩散函数(PSF)斯特列尔比,研究和比较植入不同高端人工晶状体后的临床光学图像质量。

方法

本研究纳入了194只接受以下植入的眼睛:(a)19只植入AcrySof SA60AT(对照组);(b)19只植入Miniwell;(c)24只植入LENTIS Mplus LS-313 MF30;(d)33只植入LENTIS Mplus LS-313 MF15;(e)17只植入AkkoLens Lumina;(f)31只植入AT LISA Tri 839MP;(g)20只植入Precizon Presbyopic;(h)20只植入AcrySof IQ PanOptix;(i)11只植入Tecnis Eyhance。主要观察指标为通过PWS像差仪测量的PSF斯特列尔比、排除二阶像差后的PSF斯特列尔比(PSFw2)、总均方根(RMS)、低阶像差(LOA)和高阶像差(HOA)RMS。

结果

在3.0毫米和4.0毫米瞳孔尺寸下,AT LISA Tri的PSFw2斯特列尔比最高(分别为0.52±0.14和0.31±0.10;P<0.05),其次是SA60AT(0.41±0.11和0.28±0.07)和PanOptix(0.4±0.07和0.26±0.04)。发现在3.0毫米(P<0.0001)和4.0毫米(P=0.004)时,AT LISA Tri提供的视网膜图像质量明显优于PanOptix。发现在3.0毫米(P<0.0001)和4.0毫米(P=0.002)时,Mplus MF15明显优于Mplus MF30。在所研究的组之间,总RMS、LOA RMS、HOA RMS、PSF斯特列尔比和PSFw2有显著差异(P<0.001)。

结论

通过PWS像差仪测量的远距离临床图像质量参数因植入晶状体的技术不同而有显著差异。AT LISA Tri、SA60AT和PanOptix显示出远距离视网膜图像质量的最高值,而Miniwell、Mplus MF30和Precizon Presbyopic的PSFw2斯特列尔比最低。