Suppr超能文献

针对单病灶立体定向放射治疗(SRS)和立体定向体部放射治疗(SBRT),二维和三维二极管阵列之间的患者特异性质量保证比较。

A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments.

作者信息

Lee Yongsook C, Kim Yongbok

机构信息

Department of Radiation Oncology, Miami Cancer Institute, Miami, FL 33176, USA.

Department of Radiation Oncology, Duke University, Durham, NC 27710, USA.

出版信息

J Radiosurg SBRT. 2021;7(4):295-307.

Abstract

The purpose of this study is to compare patient-specific quality assurance (PSQA) results between two dimensional (2D) diode (SRS MapCHECK) and 3D diode (ArcCHECK) arrays. Twenty-eight intracranial stereotactic radiosurgery (SRS) and 26 lung stereotactic body radiation therapy (SBRT) clinical plans with a single lesion were selected and categorized into 4 groups: 20 SRS dynamic conformal arc therapy (DCAT) plans (Group A), 8 SRS volumetric modulated arc therapy (VMAT) plans (Group B), 6 SBRT DCAT plans (Group C) and 20 SBRT VMAT plans (Group D). An individual field of each plan was delivered on SRS MapCHECK and ArcCHECK and QA analysis was performed using 4 gamma criteria of dose difference/distance-to-agreement of 3%/3 mm, 3%/2 mm, 2%/2 mm and 2%/1 mm. Statistical analysis was performed to compare PSQA results between the 2 QA devices. For all 4 groups and all 4 gamma criteria, average gamma passing rates were higher with SRS MapCHECK.

摘要

本研究的目的是比较二维(2D)二极管(SRS MapCHECK)和三维(3D)二极管(ArcCHECK)阵列之间的患者特异性质量保证(PSQA)结果。选择了28例颅内立体定向放射治疗(SRS)和26例单病灶肺部立体定向体部放射治疗(SBRT)临床计划,并分为4组:20例SRS动态适形弧形治疗(DCAT)计划(A组)、8例SRS容积调强弧形治疗(VMAT)计划(B组)、6例SBRT DCAT计划(C组)和20例SBRT VMAT计划(D组)。每个计划的单个射野在SRS MapCHECK和ArcCHECK上进行照射,并使用剂量差异/距离一致性的4种伽马标准(3%/3 mm、3%/2 mm、2%/2 mm和2%/1 mm)进行质量保证分析。进行统计分析以比较两种质量保证设备之间的PSQA结果。对于所有4组和所有4种伽马标准,SRS MapCHECK的平均伽马通过率更高。

相似文献

3
Evaluation of commercial devices for patient specific QA of stereotactic radiotherapy plans.
J Appl Clin Med Phys. 2023 Aug;24(8):e14009. doi: 10.1002/acm2.14009. Epub 2023 May 9.
6
A hybrid method to improve efficiency of patient specific SRS and SBRT QA using 3D secondary dose verification.
J Appl Clin Med Phys. 2023 Mar;24(3):e13858. doi: 10.1002/acm2.13858. Epub 2022 Dec 30.
7
Efficacious patient-specific QA for Vertebra SBRT using a high-resolution detector array SRS MapCHECK: AAPM TG-218 analysis.
J Appl Clin Med Phys. 2024 Jun;25(6):e14276. doi: 10.1002/acm2.14276. Epub 2024 Feb 27.
9
Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.
J Appl Clin Med Phys. 2024 Apr;25(4):e14258. doi: 10.1002/acm2.14258. Epub 2024 Jan 4.

引用本文的文献

4
Effect of modulation factor and low dose threshold level on gamma pass rates of single isocenter multi-target SRT treatment plans.
J Appl Clin Med Phys. 2024 Sep;25(9):e14459. doi: 10.1002/acm2.14459. Epub 2024 Jul 25.
5
Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.
J Appl Clin Med Phys. 2024 Apr;25(4):e14258. doi: 10.1002/acm2.14258. Epub 2024 Jan 4.
6
Efficient quality assurance for isocentric stability in stereotactic body radiation therapy using machine learning.
Radiol Phys Technol. 2024 Mar;17(1):219-229. doi: 10.1007/s12194-023-00768-5. Epub 2023 Dec 31.
7
A hybrid method to improve efficiency of patient specific SRS and SBRT QA using 3D secondary dose verification.
J Appl Clin Med Phys. 2023 Mar;24(3):e13858. doi: 10.1002/acm2.13858. Epub 2022 Dec 30.

本文引用的文献

1
Error sensitivity of a log file analysis tool compared with a helical diode array dosimeter for VMAT delivery quality assurance.
J Appl Clin Med Phys. 2020 Nov;21(11):163-171. doi: 10.1002/acm2.13051. Epub 2020 Oct 23.
4
Comprehensive evaluation of the high-resolution diode array for SRS dosimetry.
J Appl Clin Med Phys. 2019 Oct;20(10):13-23. doi: 10.1002/acm2.12696. Epub 2019 Sep 3.
6
Comparison of 3D and 2D gamma passing rate criteria for detection sensitivity to IMRT delivery errors.
J Appl Clin Med Phys. 2018 Jul;19(4):230-238. doi: 10.1002/acm2.12389. Epub 2018 Jun 15.
9
How the detector resolution affects the clinical significance of SBRT pre-treatment quality assurance results.
Phys Med. 2018 May;49:129-134. doi: 10.1016/j.ejmp.2017.11.012. Epub 2017 Dec 6.
10
AAPM-RSS Medical Physics Practice Guideline 9.a. for SRS-SBRT.
J Appl Clin Med Phys. 2017 Sep;18(5):10-21. doi: 10.1002/acm2.12146. Epub 2017 Aug 8.