应用回弹眼压计与压平眼压计测量晶状体不稳定犬的眼压比较:66 例(2012-2018 年)。

Comparison of intraocular pressures estimated by rebound and applanation tonometry in dogs with lens instability: 66 cases (2012-2018).

出版信息

J Am Vet Med Assoc. 2021 Nov 1;259(9):1025-1031. doi: 10.2460/javma.259.9.1025.

Abstract

OBJECTIVE

To compare intraocular pressures (IOPs) estimated by rebound and applanation tonometry for dogs with lens instability.

ANIMALS

66 dogs.

PROCEDURES

Medical records of dogs examined between September 2012 and July 2018 were reviewed for diagnoses of anterior (ALL) or posterior (PLL) lens luxation or lens subluxation.

RESULTS

Estimates of IOP obtained with rebound and applanation tonometry significantly differed from each other for all types of lens instability considered collectively (mean ± SE difference between tonometric readings, 8.1 ± 1.3 mm Hg) and specific types of lens instability considered individually (mean ± SE difference between tonometric readings: ALL, 12.8 ± 2.5 mm Hg; PLL, 5.9 ± 1.7 mm Hg; subluxation, 2.8 ± 0.8 mm Hg). Median (range) differences between rebound and applanation tonometer readings for dogs with ALL was 5 mm Hg (-9 to 76 mm Hg), with PLL was 3 mm Hg (-1 to 19 mm Hg), and with lens subluxation was 3 mm Hg (-9 to 18 mm Hg). In eyes with ALL, rebound tonometer readings exceeded applanation tonometer readings on 44 of 60 (73%) occasions.

CONCLUSIONS AND CLINICAL RELEVANCE

Rebound tonometry yielded higher estimates of IOP than did applanation tonometry in eyes with ALL and with all types of lens luxation considered collectively. Estimates of IOP in eyes with lens instability should ideally be obtained with both rebound and applanation tonometers. Veterinarians with only one type of tonometer should interpret results for dogs with lens instability concurrent with physical examination findings.

摘要

目的

比较应用回弹式眼压计和压平眼压计测量患有晶状体不稳定的犬的眼内压(IOP)。

动物

66 只狗。

程序

回顾了 2012 年 9 月至 2018 年 7 月间检查的被诊断为前(ALL)或后(PLL)晶状体脱位或晶状体半脱位的犬的病历。

结果

对于所有考虑的晶状体不稳定类型(2 种眼压计读数之间的平均差异±SE,8.1±1.3mmHg)和特定类型的晶状体不稳定(2 种眼压计读数之间的平均差异±SE:ALL,12.8±2.5mmHg;PLL,5.9±1.7mmHg;半脱位,2.8±0.8mmHg),回弹式眼压计和压平眼压计测量的 IOP 估计值之间存在显著差异。ALL 犬的回弹式眼压计和压平眼压计读数之间的中位数(范围)差异为 5mmHg(-9 至 76mmHg),PLL 为 3mmHg(-1 至 19mmHg),晶状体半脱位为 3mmHg(-9 至 18mmHg)。在 ALL 眼中,回弹眼压计读数超过压平眼压计读数的次数为 60 次中的 44 次(73%)。

结论和临床相关性

在 ALL 眼和所有类型的晶状体脱位中,回弹眼压计测量的 IOP 高于压平眼压计。晶状体不稳定眼的 IOP 估计值理想情况下应同时使用回弹眼压计和压平眼压计获得。只有一种眼压计的兽医应在与体格检查结果同时解释晶状体不稳定犬的结果。