Suppr超能文献

用于测量犬和马眼压的回弹式眼压计的评估。

Evaluation of a rebound tonometer for measuring intraocular pressure in dogs and horses.

作者信息

Knollinger Amy M, La Croix Noelle C, Barrett Paul M, Miller Paul E

机构信息

Eye Care for Animals, 1892 E Ft Union Blvd, Salt Lake City, UT 84121, USA.

出版信息

J Am Vet Med Assoc. 2005 Jul 15;227(2):244-8. doi: 10.2460/javma.2005.227.244.

Abstract

OBJECTIVE

To compare intraocular pressure (IOP) measurements obtained with a rebound tonometer in dogs and horses with values obtained by means of applanation tonometry and direct manometry.

DESIGN

Prospective study.

ANIMALS

100 dogs and 35 horses with clinically normal eyes, 10 enucleated eyes from 5 dogs, and 6 enucleated eyes from 3 horses.

PROCEDURES

In the enucleated eyes, IOP measured by means of direct manometry was sequentially increased from 5 to 80 mm Hg, and IOP was measured with the rebound tonometer. In the dogs and horses, results of rebound tonometry were compared with results of applanation tonometry.

RESULTS

For the enucleated dog and horse eyes, there was a strong (r2 = 0.99) linear relationship between pressures obtained by means of direct manometry and those obtained by means of rebound tonometry. Mean +/- SD IOPs obtained with the rebound tonometer were 10.8 +/- 3.1 mm Hg (range, 5 to 17 mm Hg) and 22.1 +/- 5.9 mm Hg (range, 10 to 34 mm Hg) for the dogs and horses, respectively. Mean IOPs obtained with the applanation tonometer were 12.9 +/- 2.7 mm Hg (range, 8 to 18 mm Hg) and 21.0 +/- 5.9 mm Hg (range, 9 to 33 mm Hg), respectively. Values obtained with the rebound tonometer were, on average, 2 mm Hg lower in the dogs and 1 mm Hg higher in the horses, compared with values obtained with the applanation tonometer.

CONCLUSIONS AND CLINICAL RELEVANCE

Results suggest that the rebound tonometer provides accurate estimates of IOP in clinically normal eyes in dogs and horses.

摘要

目的

比较用回弹式眼压计测量犬和马的眼内压(IOP)结果与用压平式眼压计和直接测压法获得的值。

设计

前瞻性研究。

动物

100只临床眼部正常的犬、35匹临床眼部正常的马、来自5只犬的10只摘除眼球以及来自3匹马的6只摘除眼球。

步骤

在摘除眼球中,通过直接测压法测量的眼内压从5毫米汞柱依次增加到80毫米汞柱,并用回弹式眼压计测量眼内压。在犬和马中,将回弹式眼压计的测量结果与压平式眼压计的结果进行比较。

结果

对于摘除眼球的犬和马的眼睛,通过直接测压法获得的压力与通过回弹式眼压计获得的压力之间存在很强的线性关系(r2 = 0.99)。犬和马用回弹式眼压计获得的平均±标准差眼内压分别为10.8±3.1毫米汞柱(范围为5至17毫米汞柱)和22.1±5.9毫米汞柱(范围为10至34毫米汞柱)。用压平式眼压计获得的平均眼内压分别为12.9±2.7毫米汞柱(范围为8至18毫米汞柱)和21.0±5.9毫米汞柱(范围为9至33毫米汞柱)。与用压平式眼压计获得的值相比,用回弹式眼压计获得的值在犬中平均低2毫米汞柱,在马中平均高1毫米汞柱。

结论及临床意义

结果表明,回弹式眼压计可为犬和马临床正常眼睛的眼内压提供准确估计。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验