Knollinger Amy M, La Croix Noelle C, Barrett Paul M, Miller Paul E
Eye Care for Animals, 1892 E Ft Union Blvd, Salt Lake City, UT 84121, USA.
J Am Vet Med Assoc. 2005 Jul 15;227(2):244-8. doi: 10.2460/javma.2005.227.244.
To compare intraocular pressure (IOP) measurements obtained with a rebound tonometer in dogs and horses with values obtained by means of applanation tonometry and direct manometry.
Prospective study.
100 dogs and 35 horses with clinically normal eyes, 10 enucleated eyes from 5 dogs, and 6 enucleated eyes from 3 horses.
In the enucleated eyes, IOP measured by means of direct manometry was sequentially increased from 5 to 80 mm Hg, and IOP was measured with the rebound tonometer. In the dogs and horses, results of rebound tonometry were compared with results of applanation tonometry.
For the enucleated dog and horse eyes, there was a strong (r2 = 0.99) linear relationship between pressures obtained by means of direct manometry and those obtained by means of rebound tonometry. Mean +/- SD IOPs obtained with the rebound tonometer were 10.8 +/- 3.1 mm Hg (range, 5 to 17 mm Hg) and 22.1 +/- 5.9 mm Hg (range, 10 to 34 mm Hg) for the dogs and horses, respectively. Mean IOPs obtained with the applanation tonometer were 12.9 +/- 2.7 mm Hg (range, 8 to 18 mm Hg) and 21.0 +/- 5.9 mm Hg (range, 9 to 33 mm Hg), respectively. Values obtained with the rebound tonometer were, on average, 2 mm Hg lower in the dogs and 1 mm Hg higher in the horses, compared with values obtained with the applanation tonometer.
Results suggest that the rebound tonometer provides accurate estimates of IOP in clinically normal eyes in dogs and horses.