Tseng Yu-Chuan, Chen Han-Sheng, Hsiao Szu-Yu, Hsu Kun-Jung, Chen Chun-Ming
School of Dentistry, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung 807, Taiwan.
Division of Orthodontics, Department of Dentistry, Kaohsiung Medical University Hospital, Kaohsiung 807, Taiwan.
Materials (Basel). 2021 Oct 12;14(20):5991. doi: 10.3390/ma14205991.
The microthread neck concept has been applied to dental implants. This study investigated the pullout strength and destruction volume of orthodontic microimplants with and without the microthread neck design. Fifteen microimplants (diameter: 1.5 × 10 mm) of three types (Types A and B: without microimplant neck; Type C: with microimplant neck) were tested. The insertion torque (IT), Periotest value (PTV), horizontal pullout strength (HPS), and horizontal destruction volume (HDV) of each type were measured. Kruskal-Wallis H test and Dunn's post-hoc comparison test were performed to compare the measured values of the three types of microimplants. The correlations of the measured values were used to perform the Spearman's correlation coefficient analysis. The ITs of Types B (8.8 Ncm) and C (8.9 Ncm) were significantly higher than those of Type A (5.2 Ncm). Type B yielded the lowest PTV (4.1), and no statistical differences in PTV were observed among the three types. Type A had a significantly lower HPS (158.8 Ncm) than Types B (226.9 Ncm) and C (212.8 Ncm). The three types did not exhibit any significant differences in the HDV. The results of the Spearman's correlation coefficient test revealed that HDV (ρ = 0.710) and IT (ρ = 0.813) were strongly correlated with HPS, whereas for PTV and HPS, it was not. HPS was strongly and significantly correlated with HDV. The orthodontic microimplant with a microimplant neck design did not perform better than that without a microthread in the mechanical strength test.
微螺纹颈部概念已应用于牙种植体。本研究调查了有无微螺纹颈部设计的正畸微种植体的拔出强度和破坏体积。测试了三种类型的15枚微种植体(直径:1.5×10mm)(A 型和 B 型:无微种植体颈部;C 型:有微种植体颈部)。测量了每种类型的植入扭矩(IT)、牙周探测值(PTV)、水平拔出强度(HPS)和水平破坏体积(HDV)。进行 Kruskal-Wallis H 检验和 Dunn 事后比较检验以比较三种类型微种植体的测量值。利用测量值的相关性进行 Spearman 相关系数分析。B 型(8.8Ncm)和 C 型(8.9Ncm)的植入扭矩显著高于 A 型(5.2Ncm)。B 型的牙周探测值最低(4.1),三种类型之间的牙周探测值未观察到统计学差异。A 型的水平拔出强度(158.8Ncm)显著低于 B 型(226.9Ncm)和 C 型(212.8Ncm)。三种类型在水平破坏体积方面未表现出任何显著差异。Spearman 相关系数检验结果显示,水平破坏体积(ρ = 0.710)和植入扭矩(ρ = 0.813)与水平拔出强度密切相关,而牙周探测值与水平拔出强度无关。水平拔出强度与水平破坏体积密切且显著相关。在机械强度测试中,有微种植体颈部设计的正畸微种植体在性能上并不优于无微螺纹的微种植体。