Optimal asymmetric thread for orthodontic microimplants: Laboratory and clinical evaluation.

作者信息

Alrbata Raed H, Ha Do-Won, Yu Wonjae, Kyung Hee-Moon

机构信息

a  Former Research Assistant, Department of Orthodontics, School of Dentistry, Kyungpook National University, Daegu, Korea.

b  Orthodontist, Mir Dental Hospital, Daegu, Korea.

出版信息

Angle Orthod. 2015 Jul;85(4):585-90. doi: 10.2319/050314-320.1. Epub 2014 Oct 6.

Abstract

OBJECTIVES

To investigate the performance of microimplants incorporating a newly designed asymmetric thread.

MATERIALS AND METHODS

Three microimplants were compared. The control group comprised microimplants with the original v-shaped thread. The two experimental groups (Taper 1.0 and Taper 1.25) comprised prototype microimplants constructed with the new asymmetric thread; the Taper 1.25 specimens had a 1.25-mm-long and sharper tip, while the Taper 1.0 and control groups had a less sharp 1-mm tip. Two specially designed artificial bone blocks mimicking soft (maxillary) and hard (mandibular) bone were used to evaluate the microimplant insertion characteristics and postinsertion lateral stability. The peak insertion torque, insertion time, Periotest value (PTV), and torsional strength were measured. Then the microimplants were evaluated clinically over a 3-month period.

RESULTS

Significant differences in peak insertion torque, insertion time, and PTV were observed and favored the experimental groups. Although statistically insignificant, the clinical success rate was also higher in the Taper 1.25 experimental group than in the control group (87.2% vs 75.6%).

CONCLUSIONS

The better performances of the experimental microimplant, under both laboratory and clinical conditions (although statistically insignificant in the latter), demonstrate the superiority of the new asymmetric thread.

摘要

目的

研究采用新设计的不对称螺纹的微型种植体的性能。

材料与方法

比较三种微型种植体。对照组为具有原始V形螺纹的微型种植体。两个实验组(锥度1.0和锥度1.25)为由新型不对称螺纹构建的原型微型种植体;锥度1.25的样本有一个1.25毫米长且更尖锐的尖端,而锥度1.0组和对照组有一个不太尖锐的1毫米尖端。使用两个专门设计的模拟软质(上颌)和硬质(下颌)骨的人工骨块来评估微型种植体的植入特性和植入后的侧向稳定性。测量峰值植入扭矩、植入时间、牙周探测值(PTV)和抗扭强度。然后在3个月的时间内对微型种植体进行临床评估。

结果

观察到峰值植入扭矩、植入时间和PTV存在显著差异,且有利于实验组。虽然在统计学上不显著,但锥度1.25实验组的临床成功率也高于对照组(87.2%对75.6%)。

结论

实验性微型种植体在实验室和临床条件下均表现出更好的性能(尽管后者在统计学上不显著),证明了新型不对称螺纹的优越性。

相似文献

8
Potential of self-drilling orthodontic microimplants under immediate loading.自攻式正畸微种植体即刻加载的潜力。
Am J Orthod Dentofacial Orthop. 2010 Apr;137(4):496-502. doi: 10.1016/j.ajodo.2008.06.030.

本文引用的文献

1
Effects of thread shape on the pullout strength of miniscrews.螺纹形状对微螺钉拔出强度的影响。
Am J Orthod Dentofacial Orthop. 2012 Aug;142(2):186-90. doi: 10.1016/j.ajodo.2012.03.023.
4
Cortical bone thickness at common miniscrew implant placement sites.常见微型种植体植入部位皮质骨厚度。
Am J Orthod Dentofacial Orthop. 2011 Apr;139(4):495-503. doi: 10.1016/j.ajodo.2009.03.057.
6
Miniscrew stability evaluated with computerized tomography scanning.应用计算机断层扫描评价微螺钉稳定性。
Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):73-9. doi: 10.1016/j.ajodo.2008.03.024.
8
10
Survival analyses of surgical miniscrews as orthodontic anchorage.作为正畸支抗的微型外科螺钉的生存分析。
Am J Orthod Dentofacial Orthop. 2009 Jul;136(1):29-36. doi: 10.1016/j.ajodo.2007.06.018.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验