Alrbata Raed H, Ha Do-Won, Yu Wonjae, Kyung Hee-Moon
a Former Research Assistant, Department of Orthodontics, School of Dentistry, Kyungpook National University, Daegu, Korea.
b Orthodontist, Mir Dental Hospital, Daegu, Korea.
Angle Orthod. 2015 Jul;85(4):585-90. doi: 10.2319/050314-320.1. Epub 2014 Oct 6.
To investigate the performance of microimplants incorporating a newly designed asymmetric thread.
Three microimplants were compared. The control group comprised microimplants with the original v-shaped thread. The two experimental groups (Taper 1.0 and Taper 1.25) comprised prototype microimplants constructed with the new asymmetric thread; the Taper 1.25 specimens had a 1.25-mm-long and sharper tip, while the Taper 1.0 and control groups had a less sharp 1-mm tip. Two specially designed artificial bone blocks mimicking soft (maxillary) and hard (mandibular) bone were used to evaluate the microimplant insertion characteristics and postinsertion lateral stability. The peak insertion torque, insertion time, Periotest value (PTV), and torsional strength were measured. Then the microimplants were evaluated clinically over a 3-month period.
Significant differences in peak insertion torque, insertion time, and PTV were observed and favored the experimental groups. Although statistically insignificant, the clinical success rate was also higher in the Taper 1.25 experimental group than in the control group (87.2% vs 75.6%).
The better performances of the experimental microimplant, under both laboratory and clinical conditions (although statistically insignificant in the latter), demonstrate the superiority of the new asymmetric thread.
研究采用新设计的不对称螺纹的微型种植体的性能。
比较三种微型种植体。对照组为具有原始V形螺纹的微型种植体。两个实验组(锥度1.0和锥度1.25)为由新型不对称螺纹构建的原型微型种植体;锥度1.25的样本有一个1.25毫米长且更尖锐的尖端,而锥度1.0组和对照组有一个不太尖锐的1毫米尖端。使用两个专门设计的模拟软质(上颌)和硬质(下颌)骨的人工骨块来评估微型种植体的植入特性和植入后的侧向稳定性。测量峰值植入扭矩、植入时间、牙周探测值(PTV)和抗扭强度。然后在3个月的时间内对微型种植体进行临床评估。
观察到峰值植入扭矩、植入时间和PTV存在显著差异,且有利于实验组。虽然在统计学上不显著,但锥度1.25实验组的临床成功率也高于对照组(87.2%对75.6%)。
实验性微型种植体在实验室和临床条件下均表现出更好的性能(尽管后者在统计学上不显著),证明了新型不对称螺纹的优越性。