皮肤镜磁体相关的电磁场可能影响心脏植入式电子设备患者的安全。

Electromagnetic Field Associated With Dermoscope Magnets May Affect the Safety of Cardiac Implanted Electronic Devices Patients.

作者信息

Sławiński Grzegorz, Sławińska Martyna, Usarek Zbigniew, Sobjanek Michał, Kempa Maciej, Liżewska-Springer Aleksandra, Lewicka Ewa, Nowicki Roman J, Raczak Grzegorz

机构信息

Department of Cardiology and Electrotherapy, Faculty of Medicine, Medical University of Gdańsk, Gdańsk, Poland.

Department of Dermatology, Venereology and Allergology, Faculty of Medicine, Medical University of Gdańsk, Gdańsk, Poland.

出版信息

Front Cardiovasc Med. 2021 Oct 14;8:757032. doi: 10.3389/fcvm.2021.757032. eCollection 2021.

Abstract

Dermoscopy is currently used as an auxiliary tool in general dermatology. Since some commercially available dermoscopes have built-in magnets, electromagnetic interference (EMI) may occur when examining cardiac implantable electronic devices (CIED) patients. The aim of the study was to create maps of electromagnetic fields defining a safe distance in terms of EMI. The study was performed in laboratory conditions using measuring equipment specially designed for this purpose. The following dermoscopes have been tested: Illuco IDS-1100, Visiomed Luminis, Visiomed Luminis 2, Heine NC2 with and without a contact plate, DermLite DL4, and DermLite Handyscope. Measurements were made for the following set of lift-off distances: 5, 10, 20, 30, 40, 50, and 150 mm. Each 2D scan consisted of 10-line scans shifted from each other by 10 mm. The strength of the magnetic field decreased with the distance from the faceplate. The distribution of the magnetic field differed depending on the position of the magnets. The highest magnetic field was recorded in the center of the Heine NC2 faceplate (up to 8 mT). In most cases, at a distance of 10 mm, the magnetic field strength was measured below 1 mT, with the exception of Heine NC2 and Heine NC2 with a contact plate. All tested dermoscopes generated a magnetic field of <1 mT at the distance of 20 mm. The use of dermoscopes with built-in magnets may affect the functioning of CIEDs, and the impact may vary depending on the type of dermoscope.

摘要

皮肤镜目前在普通皮肤科中用作辅助工具。由于一些市售的皮肤镜内置有磁铁,在检查心脏植入式电子设备(CIED)患者时可能会发生电磁干扰(EMI)。本研究的目的是绘制电磁场图,确定在电磁干扰方面的安全距离。该研究在实验室条件下使用为此专门设计的测量设备进行。测试了以下皮肤镜:Illuco IDS - 1100、Visiomed Luminis、Visiomed Luminis 2、带和不带接触板的Heine NC2、DermLite DL4以及DermLite Handyscope。针对以下一组提离距离进行了测量:5、10、20、30、40、50和150毫米。每次二维扫描由10条线扫描组成,彼此偏移10毫米。磁场强度随着与面板距离的增加而降低。磁场分布因磁铁位置而异。Heine NC2面板中心记录到的磁场最强(高达8毫特斯拉)。在大多数情况下,在10毫米的距离处,除了Heine NC2和带接触板的Heine NC2外,磁场强度测量值低于1毫特斯拉。所有测试的皮肤镜在20毫米的距离处产生的磁场均<1毫特斯拉。使用内置磁铁的皮肤镜可能会影响CIED的功能,并且影响可能因皮肤镜类型而异。