Wear rate and osteolysis in two types of second-generation annealed highly cross-linked polyethylene in total hip arthroplasty: A retrospective comparative study with a minimum of five years.

作者信息

Yamamoto Naoki, Takada Ryohei, Jinno Tetsuya, Miyatake Kazumasa, Watanabe Naoto, Koga Hideyuki, Yoshii Toshitaka, Yagishita Kazuyoshi, Okawa Atsushi

机构信息

Department of Orthopaedic Surgery Medical Hospital, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, 113-8519 Tokyo, Japan.

Department of Orthopaedic Surgery Medical Hospital, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, 113-8519 Tokyo, Japan.

出版信息

Orthop Traumatol Surg Res. 2023 Feb;109(1):103147. doi: 10.1016/j.otsr.2021.103147. Epub 2021 Nov 15.

Abstract

BACKGROUND

As no previous study has directly compared the linear wear rate in two types of second-generation annealed highly cross-linked polyethylene, we performed a retrospective study with a minimum of 5-year follow-up to assess primary arthroplasties in the (1) wear rates and (2) incidence of osteolysis of the two types of HXLPE.

HYPOTHESIS

There was no significant difference in the linear wear rate and the incidence of osteolysis between the two types of second-generation annealed highly cross-linked polyethylene.

PATIENTS AND METHODS

In this single-center study, we reviewed 257 cases of primary cementless total hip arthroplasties between 2011 and 2015, which were performed with 32mm delta ceramic on second-generation annealed highly cross-linked polyethylene (X3 and E1 were used in 105 and 103 cases, respectively.). The mean wear rate was evaluated using a computer-assisted method, and the incidence of osteolysis was evaluated based on the appearance of a localized area with loss of trabecular bone or cortical erosion adjacent to the implants during the latest follow-up.

RESULTS

In total, we evaluated 208 cases, followed postoperatively for over 5 years (mean, 6.1 years, range: 5.0-8.0). There were no significant differences between the two groups with respect to age (list in order of Group X, Group E, p value) (61.2±12.3, 62.7±12.1, p=0.36), sex (ratio of male: 17.1%, 14.6%, p=0.61), body mass index (22.9±3.7, 22.8±4.0, p=0.91), pre- (49.9±14.8, 48.5±13.8, p=0.49) and post-operative (91.3±9.1, 92.7±7.0, p=0.23) Japanese Orthopaedic Association Hip Score, cup size (50.8±3.0, 50.9±2.2, p=0.70), cup inclination (38.7±4.8, 37.6±4.8, p=0.10), and cup anteversion (18.7±6.9, 18.5±7.6, p=0.80). The mean linear wear rates of the X3 and E1 groups were 0.057±0.039 (range: 0-0.16) and 0.054±0.037mm/year (range: 0-0.15), respectively (p=0.61). No osteolysis was found on the final plain radiographs in both groups.

DISCUSSION

This study revealed that both types of highly cross-linked polyethylene have excellent linear wear rates and were equally safe to use. However, the difference between the two materials in terms of the long-term wear rate should be further validated.

LEVEL OF EVIDENCE

III; retrospective case control study.