Evidence for reduced torsional resistance of rotary files under curved position.

作者信息

Jamleh Ahmed, Almedlej Raghad, Alomar Rasha, Almayouf Nouf, Alfadley Abdulmohsen, Alfouzan Khalid

机构信息

Restorative and Prosthetic Dental Sciences, College of Dentistry, King Saud bin Abdulaziz University for Health Sciences, National Guard Health Affairs, Riyadh, Saudi Arabia.

King Abdullah International Medical Research Centre, National Guard Health Affairs, Riyadh, Saudi Arabia.

出版信息

Saudi Dent J. 2021 Nov;33(7):614-619. doi: 10.1016/j.sdentj.2020.07.005. Epub 2020 Aug 6.

Abstract

OBJECTIVES

The aim of this in vitro study was to investigate the torsional fatigue resistance of different rotary files in straight and curved positions at a simulated intracanal temperature.

METHODS

Forty-eight size #25 files from ProTaper Universal (PTU), ProTaper Next (PTN) and ProTaper Gold (PTG) were tested for torsional resistance in both straight and curved (90 degrees) positions (n = 8 each). The experiment was conducted at a simulated intracanal temperature of 35 ± 1 °C. The torque gauge was reset before every use, and uniform torsional stress was applied by rotating the file clockwise at a speed of 40 rpm until fracture. Torque failure values were recorded, and fractured surfaces were tested using a scanning electron microscopy (SEM). Statistical analysis of the data was completed using one-way ANOVA and post-hoc test (LSD) in order to compare between the tested systems in each file position. Student's -test was also performed to compare between the two positions in each rotary system, and significance level was set at 5%.

RESULTS

In the straight position, torsional resistance of PTN was significantly lower than that of PTU and PTG (p < 0.001), while the PTU and PTG had comparable results (p > 0.05). In the curved position, torsional resistance was found to be comparable among all systems (p > 0.05). Within each system, the curved position resulted in a significantly lower torsional resistance when compared to the straight position (p < 0.001). SEM examination revealed similar torsional failure behavior for all files tested under both positions.

CONCLUSIONS

Our study identified that placing files in a curved position was associated with a significant decrease in torsional resistance of all systems, and this necessitates reconsidering the torque settings proposed by the manufacturers. Whether these effects can be translated into in vivo studies remains to be investigated, and thus future studies are essential to provide confirming evidence.

摘要

目的

本体外研究旨在调查不同旋转锉在模拟根管内温度下直线和弯曲位置的抗扭转疲劳性能。

方法

对来自ProTaper Universal(PTU)、ProTaper Next(PTN)和ProTaper Gold(PTG)的48根#25号锉进行直线和弯曲(90度)位置的抗扭转测试(每组n = 8)。实验在35 ± 1 °C的模拟根管内温度下进行。每次使用前重新设置扭矩计,并以40 rpm的速度顺时针旋转锉直至折断,施加均匀的扭转应力。记录扭矩失效值,并使用扫描电子显微镜(SEM)对断口表面进行检测。使用单因素方差分析和事后检验(LSD)对数据进行统计分析,以比较每个锉位置的测试系统。还进行了学生t检验以比较每个旋转系统的两个位置,显著性水平设定为5%。

结果

在直线位置,PTN的抗扭转性能显著低于PTU和PTG(p < 0.001),而PTU和PTG的结果相当(p > 0.05)。在弯曲位置,所有系统的抗扭转性能相当(p > 0.05)。在每个系统中,与直线位置相比,弯曲位置导致抗扭转性能显著降低(p < 0.001)。SEM检查显示,在两个位置测试的所有锉的扭转失效行为相似。

结论

我们的研究发现,将锉置于弯曲位置会导致所有系统的抗扭转性能显著下降,这需要重新考虑制造商建议的扭矩设置。这些影响是否能转化为体内研究仍有待调查,因此未来的研究对于提供确凿证据至关重要。