Appropriate soft tissue tension in total knee replacement (TKR) is an important factor for a successful outcome. The purpose of our study was to assess both the reproducibility of a modern intraoperative pressure sensor (IOP) and if a surgeon could unconsciously influence measurement. A consecutive series of 80 TKRs were assessed with an IOP between January 2018 and December 2020. In the first scenario, two blinded sequential measurements in 48 patients were taken; in a second scenario, an initial blinded measurement and a subsequent unblinded measurement in 32 patients were taken while looking at the sensor monitor screen. Reproducibility was assessed by intraclass correlation coefficients (ICCs). In the first scenario, the ICC ranged from 0.83 to 0.90, and in the second scenario it ranged from 0.80 to 0.90. All ICCs were 0.80 or higher, indicating reproducibility using a IOP and that a surgeon may not unconsciously influence the measurement. The use of a modern IOP to measure soft tissue tension in TKRs is a reproducible technique. A surgeon observing the measurements while performing IOP may not significantly influence the result. An IOP gives additional information that the surgeon can use to optimize outcomes in TKR.
全膝关节置换术(TKR)中适当的软组织张力是手术成功的重要因素。我们研究的目的是评估一种现代术中压力传感器(IOP)的可重复性,以及外科医生是否会在无意识的情况下影响测量结果。在2018年1月至2020年12月期间,对连续的80例TKR患者使用IOP进行评估。在第一种情况下,对48例患者进行了两次盲法连续测量;在第二种情况下,对32例患者进行了一次初始盲法测量和随后的非盲法测量,测量时查看传感器监测屏幕。通过组内相关系数(ICC)评估可重复性。在第一种情况下,ICC范围为0.83至0.90,在第二种情况下,ICC范围为0.80至0.90。所有ICC均为0.80或更高,表明使用IOP具有可重复性,并且外科医生可能不会在无意识的情况下影响测量。使用现代IOP测量TKR中的软组织张力是一种可重复的技术。外科医生在进行IOP测量时观察测量结果可能不会显著影响结果。IOP可提供额外信息,外科医生可利用这些信息优化TKR的手术效果。