Becze Anca, Babalau Fuss Vanda Liliana, Scurtu Daniela Alexandra, Tomoaia-Cotisel Maria, Mocanu Aurora, Cadar Oana
Research Institute for Analytical Instrumentation, National Institute of Research and Development for Optoelectronics INOE 2000, 67 Donath Street, 400293 Cluj-Napoca, Romania.
Faculty of Chemistry and Chemical Engineering, Research Center of Physical Chemistry, Babeş-Bolyai University, 11 Arany Janos Street, 400028 Cluj-Napoca, Romania.
Molecules. 2021 Nov 19;26(22):6982. doi: 10.3390/molecules26226982.
The content and composition of dietary supplements is of great interest due to their increasing consumption and variety of available brand offered in the market. Accurate determination of vitamins is important for the improvement of dietary supplement quality and nutrition assessments. In this regard, the simultaneous determination of vitamin D3 (calcitriol-CT and cholecalciferol-CHL) and K2 (menaquinone-4-MK-4 and menaquinone-7-MK-7) in dietary supplements was developed by using ultra-high-pressure liquid chromatography (UHPLC). The overall runtime per sample was above 35 min, with the retention times of 2.40, 6.59, 7.06, and 32.6 min for vitamin D3 (CT and CHL) and vitamin K2 (MK-4 and MK-7), respectively. The limits of detection and limits of quantification for the target nutritional compounds ranged between 0.04-0.05 µg/mL, respectively. The validation results indicated that the method had reasonable linearity ( ≥ 0.9990), good recovery (>82%), satisfactory intra-day precision (≤1.9%) and inter-day precision (≤3.5%), and high selectivity and specificity. The validated UHPLC method was demonstrated to be precise, accurate, and robust for the simultaneous determination of vitamins D3 (CT and CHL) and K2 (MK-4 and MK-7) in dietary supplements.
由于膳食补充剂的消费量不断增加以及市场上提供的品牌种类繁多,其成分和组成备受关注。准确测定维生素对于提高膳食补充剂质量和营养评估至关重要。在这方面,采用超高压液相色谱(UHPLC)建立了同时测定膳食补充剂中维生素D3(骨化三醇-CT和胆钙化醇-CHL)和维生素K2(甲基萘醌-4-MK-4和甲基萘醌-7-MK-7)的方法。每个样品的总运行时间超过35分钟,维生素D3(CT和CHL)和维生素K2(MK-4和MK-7)的保留时间分别为2.40、6.59、7.06和32.6分钟。目标营养化合物的检测限和定量限分别在0.04 - 0.05 µg/mL之间。验证结果表明,该方法具有合理的线性(≥0.9990)、良好的回收率(>82%)、令人满意的日内精密度(≤1.9%)和日间精密度(≤3.5%),以及高选择性和特异性。经验证的UHPLC方法被证明对于同时测定膳食补充剂中的维生素D3(CT和CHL)和维生素K2(MK-4和MK-7)是精确、准确和稳健的。