Performance of the Uni-Vent Eagle™ Model 754 ventilator under hyperbaric conditions.

作者信息

Popa Daniel, Waterhouse Lynn, Duchnick Jay, Neuman Tom, Witucki Peter

机构信息

Hennepin County Medical Center, Minneapolis, Minnesota U.S.

Daniel P. Haerther Center for Conservation and Research, John G. Shedd Aquarium, Chicago, Illinois U.S.

出版信息

Undersea Hyperb Med. 2021 Fourth Quarter;48(4):409-416.

Abstract

Critically ill patients needing mechanical ventilation may require hyperbaric oxygen therapy. Some institutions still use ventilators that were available prior to the advent of hyperbaric-specific units, such as the Uni-Vent Eagle™ Model model 754. Here we examine the performance of the Uni-Vent model 754 under hyperbaric conditions and investigate concerns of an oxygen leak in the ventilator housing, which poses a fire risk. We studied the ventilator at 1.0, 2.4 and 2.8 ATA in assist control mode using a Michigan test lung and a variety of tidal volumes and respiratory rates. We recorded the delivered volumes, peak pressures, and oxygen percentages within the hyperbaric chamber at 2.4 and 2.8 ATA and within the ventilator housing. At those pressures the ventilator delivered approximately 25% less volume than at 1.0 ATA. We observed breath stacking at high respiratory rates, but this was blunted at both 2.4 and 2.8 ATA. Oxygen levels did not rise in the housing during our investigation. In addition, we fit a linear regression to the data comparing set tidal volumes and delivered tidal volumes in order to model the changes observed. Hyperbaric conditions caused decreased delivered tidal volumes in a depth-dependent fashion, and oxygen levels within the housing did not rise. The Uni-Vent Eagle model 754 performed safely and effectively at depth but requires spirometry to correctly program desired ventilator settings.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验