Adelaide Dental School, Adelaide Health & Medical Sciences, The University of Adelaide, South Australia, Australia.
Adelaide Dental School, The University of Adelaide, South Australia, Australia.
Aust Dent J. 2022 Jun;67(2):125-131. doi: 10.1111/adj.12891. Epub 2021 Dec 9.
The purpose of this study was to compare the marginal gaps of CAD/CAM lithium disilicate (LDS) crowns constructed using three different scanner and milling unit combinations.
Twenty-four Columbia model lower left molars were prepared for LDS crowns by undergraduate students in a simulated environment. One LDS crown was constructed for each crown preparation using each of the following intraoral scanner/milling unit combinations: TRIOS 3 scanner/E4D milling unit (TRIO/E4DM); TRIOS 3 scanner/Sirona inLab MC X5 milling unit (TRIO/MCX5); E4D scanner/E4D milling unit (E4DS/E4DM). The crowns were seated onto the original crown preparations and three vertical marginal gap measurements taken at four locations (mid-buccal, mid-lingual, mid-mesial and mid-distal) using a stereomicroscope. The mean marginal gap (MMG) was calculated for each crown and each individual tooth surface.
The TRIO/MCX5 scanner/milling unit combination produced the smallest MMG of 63.73 ± 47.38 μm followed by E4DS/E4DM (88.64 ± 106.51 μm) and TRIO/E4DM (107.41 ± 76.47 μm). There was a significant effect of milling unit (p < 0.0001) on MMG but no significant effect of scanner (p = 0.070) or location (p = 0.182).
The newer scanner/milling unit combination produced LDS crowns with the smallest MMG. All scanner/milling unit combinations produced LDS crowns within a clinically acceptable MMG.
本研究的目的是比较使用三种不同的扫描仪和铣削单元组合构建 CAD/CAM 锂硅(LDS)冠的边缘间隙。
在模拟环境中,由本科生为 24 个哥伦比亚模型左下磨牙制备 LDS 冠。使用以下每种口内扫描仪/铣削单元组合为每个冠制备构建一个 LDS 冠:TRIOS 3 扫描仪/E4D 铣削单元(TRIO/E4DM);TRIOS 3 扫描仪/Sirona inLab MC X5 铣削单元(TRIO/MCX5);E4D 扫描仪/E4D 铣削单元(E4DS/E4DM)。将冠放置在原始冠制备体上,并使用立体显微镜在四个位置(颊中、舌中、近中、远中)测量三个垂直边缘间隙。为每个冠和每个个体牙面计算平均边缘间隙(MMG)。
TRIO/MCX5 扫描仪/铣削单元组合产生的最小 MMG 为 63.73±47.38μm,其次是 E4DS/E4DM(88.64±106.51μm)和 TRIO/E4DM(107.41±76.47μm)。铣削单元对 MMG 有显著影响(p<0.0001),但扫描仪(p=0.070)和位置(p=0.182)无显著影响。
较新的扫描仪/铣削单元组合制造的 LDS 冠具有最小的 MMG。所有扫描仪/铣削单元组合制造的 LDS 冠均在临床可接受的 MMG 范围内。