定量宫颈应变弹性成像(E-Cervix)在妊娠期的可重复性。

Repeatability and reproducibility of quantitative cervical strain elastography (E-Cervix) in pregnancy.

机构信息

Collegium Medicum, Jan Kochanowski University in Kielce, Kielce, Poland.

Clinic of Obstetrics and Gynaecology, Provincial Combined Hospital in Kielce, Kielce, Poland.

出版信息

Sci Rep. 2021 Dec 8;11(1):23689. doi: 10.1038/s41598-021-02498-3.

Abstract

Strain elastography of the uterine cervix may be useful in the diagnosis and prediction of obstetric complications. The inability to obtain quantitative results, with only the possibility of visual semiquantitative evaluation of the obtained elastograms, has been the limitation of the method thus far. E-Cervix is a software program that uses intrinsic compression to excite tissue and allows the evaluation of quantitative parameters on the basis of pixel distribution in an elastogram. The aim of this study was to assess the repeatability and reproducibility of quantitative cervical strain elastography (E-Cervix) of the uterine cervix and to assess the correlation of the obtained parameters with selected clinical features of patients in the third trimester of pregnancy. In total, 222 patients participated in the study. We assessed 5 ultrasound parameters: elasticity index (ECI), hardness ratio (HR), internal os strain (IOS), external os strain (EOS) and IOS/EOS ratio. Each study was performed according to a predetermined standardized protocol. For all assessed elastographic parameters, we obtained good intra- and interobserver reproducibility. The interclass correlation coefficient (ICC) ranged from 0.77 to 0.838 for intraobserver variability and from 0.771 to 0.826 for interobserver variability. We demonstrated a significant correlation of some obtained elastographic parameters with the basic clinical features of patients, such as age, the number of previous caesarean sections, pregnancy weight and BMI. In each case, the correlation was very low. Quantitative elastographic assessment with the use of E-Cervix is characterized by good repeatability. Some clinical features may affect the value of the parameters obtained. The clinical relevance of this interference requires further investigation.

摘要

子宫颈应变弹性成像在产科并发症的诊断和预测中可能具有一定的作用。但目前该方法的局限性在于无法获得定量结果,只能对获得的弹性图像进行可视化半定量评估。E-Cervix 是一款软件程序,它使用固有压缩来激发组织,并允许根据弹性图像中的像素分布来评估定量参数。本研究旨在评估子宫颈应变弹性成像(E-Cervix)的重复性和再现性,并评估所获得的参数与妊娠晚期患者的选定临床特征之间的相关性。共有 222 名患者参与了本研究。我们评估了 5 个超声参数:弹性指数(ECI)、硬度比(HR)、内口应变(IOS)、外口应变(EOS)和 IOS/EOS 比。每个研究都是根据预先确定的标准化方案进行的。对于所有评估的弹性成像参数,我们都获得了良好的观察者内和观察者间可重复性。组内观察者间的 ICC 范围为 0.77 至 0.838,组间观察者间的 ICC 范围为 0.771 至 0.826。我们证明了一些获得的弹性成像参数与患者的基本临床特征之间存在显著相关性,例如年龄、既往剖宫产次数、妊娠体重和 BMI。在每种情况下,相关性都非常低。使用 E-Cervix 进行定量弹性评估具有良好的重复性。一些临床特征可能会影响所获得参数的价值。这种干扰的临床意义需要进一步研究。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a126/8655038/45c97771dc2c/41598_2021_2498_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验