Ohrloff C, Schalnus R, Spitznas M
Klin Monbl Augenheilkd. 1986 Jul;189(1):24-7. doi: 10.1055/s-2008-1050741.
Whereas specular microscopy yields mostly qualitative information, fluorophotometry furnishes quantitative data of endothelial cell function. Determination of fluorescein permeability of the endothelial cell layer reflects the function of the endothelial barrier. Following topical application of fluorescein, the time-dependent change in the fluorescein concentration in corneal stroma and aqueous yields the transfer coefficient (Kc) of the corneal endothelium. With the Fluorotron Master, Kc was 3.77 +/- 0.57 X 10(-3)/min. for normal eyes; patients with cornea guttata or Fuchs's dystrophy had a significantly higher transfer coefficient (Kc = 7.9 +/- 2.88 X 10(-3)/min.). Normal Kc values were found 6-63 months (average 33.5 months) after phacoemulsification with implantation of a posterior chamber lens.
虽然镜面显微镜检查大多产生定性信息,但荧光光度法可提供内皮细胞功能的定量数据。内皮细胞层荧光素通透性的测定反映了内皮屏障的功能。局部应用荧光素后,角膜基质和房水中荧光素浓度随时间的变化得出角膜内皮的转运系数(Kc)。使用Fluorotron Master,正常眼睛的Kc为3.77±0.57×10⁻³/分钟;角膜滴状变性或富克斯角膜内皮营养不良患者的转运系数显著更高(Kc = 7.9±2.88×10⁻³/分钟)。在超声乳化联合后房型人工晶状体植入术后6 - 63个月(平均33.5个月)发现Kc值正常。