Eur J Ophthalmol. 2022 Sep;32(5):2967-2974. doi: 10.1177/11206721211069737. Epub 2021 Dec 21.
To evaluate visual performance with trifocal and extended depth of focus IOL at 1 year post-operatively.
BMI Southend Hospital.
Cohort study.
An age-matched cohort of forty subjects bilaterally implanted with the AT LISA 839MP trifocal IOL (20 patients, 40 eyes) and the Tecnis Symfony extended depth of focus IOL (20 patients, 40 eyes) were assessed at 3-6 months and 12-18 months post-operatively. Primary outcome measures were distance (6 m), intermediate (70 cm), near visual acuity (40 cm), and analysis of defocus profiles. Secondary outcomes included contrast sensitivity, Radner reading performance, quality of vision and assessment of halos.
Distance visual acuity (VA) and defocus areas were similar (= 0.07). No significant difference in intermediate VA was noted but the intermediate area of focus was greater in the EDoF (0.31 ± 0.12 LogMARm) compared to the trifocal (0.22 ± 0.08LogMARm) ( = 0.02). However, all near metrics were significantly better in the trifocal group. 80% of trifocal subjects were spectacle independent compared to 50% EDoF subjects. Quality of vision questionnaire found no significant differences between groups, however halo scores were greater at 3-6 months in the trifocal group (< 0.01) but no differences were noted at 12-18 months.
Near vision is significantly better for the trifocal, thus greater levels of spectacle independence. The range of intermediate vision was greater for the EDoF but no difference in intermediate VA. In the early period, differences in contrast sensitivity and halo size/intensity were noted, however, by one-year these measures were not significantly different.
评估三焦点和扩展景深人工晶状体术后 1 年的视觉效果。
BMI 绍森德医院。
队列研究。
对 40 名年龄匹配的患者(20 名患者,40 只眼)分别植入 AT LISA 839MP 三焦点人工晶状体和 Tecnis Symfony 扩展景深人工晶状体,在术后 3-6 个月和 12-18 个月进行评估。主要观察指标为远视力(6m)、中距离视力(70cm)、近视力(40cm)和离焦曲线分析。次要观察指标包括对比敏感度、Radner 阅读表现、视觉质量和评估光晕。
远视力(VA)和离焦区相似(=0.07)。中距离 VA 无显著差异,但 EDoF 组的中距离焦点区域大于三焦点组(0.31±0.12 LogMARm 比 0.22±0.08LogMARm)(=0.02)。然而,所有近视力指标在三焦点组均显著更好。80%的三焦点组患者无需戴镜,而 EDoF 组为 50%。视觉质量问卷发现两组间无显著差异,但三焦点组在术后 3-6 个月时的光晕评分更高(<0.01),但在 12-18 个月时无差异。
三焦点的近视力显著更好,因此无需戴镜的比例更高。EDoF 的中距离视力范围更大,但中距离 VA 无差异。在早期,对比敏感度和光晕大小/强度存在差异,但在 1 年时这些指标无显著差异。