Chem Commun (Camb). 2022 Feb 22;58(16):2706-2709. doi: 10.1039/d1cc07115g.
A novel non-enzymatic electrochemical sensor was constructed to achieve a highly selective and ultrasensitive detection of cholesterol to address the problems related to serious coronary heart disease. The NiVP/Pi-based sensor exhibited ultra-high sensitivity of 5510.18 μA μM cm and 36.8 μA μM cm for 1 nM to 10 μM and 100 μM to 10 mM of cholesterol, respectively, with an ultra-low detection limit of 1 aM, along with superior selectivity for cholesterol when exposed to various interferents such as ascorbic acid, glucose and uric acid. Also, a novel NiVP/Pi-based flexible sensor coated onto Whatman filter paper was developed and displayed superior sensitivity, even with a human blood serum sample at physiological pH.