Suppr超能文献

数字乳腺断层合成系统中水平和垂直平面的散射辐射强度。

SCATTER RADIATION INTENSITIES IN HORIZONTAL AND VERTICAL PLANES ABOUT DIGITAL BREAST TOMOSYNTHESIS SYSTEM.

机构信息

Medical Physics Department, Austin Health, 145 Studley Road, Melbourne, Victoria 3084, Australia.

出版信息

Radiat Prot Dosimetry. 2022 Feb 18;198(1-2):119-127. doi: 10.1093/rpd/ncab197.

Abstract

Digital breast tomosynthesis (DBT) is a main imaging modality for breast imaging. However, shielding calculations for DBT are commonly based on previous technology with softer beam spectra. In addition, shielding calculations often assume some patient attenuation is provided in locations to the posterior of the patient but without quantification of this attenuation. In this work, recent research in DBT shielding is validated by measuring scatter radiation in the vertical plane. It is also extended upon by measuring the scatter radiation in the horizontal plane and the effect of patient attenuation. These measurements are weighted by our local patient compressed breast thickness distribution to provide a simple scattering factor. Air kerma distributions are provided demonstrating the distribution of scatter radiation around the DBT system in the presence of patient body attenuation. The highest air kerma measured in any direction from a 4-view per patient screening tomosynthesis exam based on the local patient workload is 55 μGy at 1 m. This value can be used to conservatively treat the scatter radiation as an isotropic distribution for shielding assessments. This work is consistent with many recent publications, with the notable exception of the forward scattered peak which is smaller than previous studies. Under our locally encountered breast thickness distribution, the increased scatter radiation and consequent increased minimum shielding requirements is small in most cases. With 400 patients per week, the air kerma at 1 m is 22 mGy and, for a fully occupied public area 2 m away, this requires a gypsum plasterboard thickness of 18.0 mm which is 3.6 mm more than previous data suggests. Although the increase is minimal, standard manufacturing thicknesses of shielding material may not be automatically assumed to be sufficient and careful consideration of the scatter radiation is warranted.

摘要

数字乳腺断层合成(DBT)是乳腺成像的主要成像方式。然而,DBT 的屏蔽计算通常基于具有更软光束光谱的先前技术。此外,屏蔽计算通常假设在患者后部的某些位置提供了一些患者衰减,但没有对这种衰减进行量化。在这项工作中,通过测量垂直平面中的散射辐射来验证 DBT 屏蔽的最新研究。还通过测量水平平面中的散射辐射和患者衰减的影响来扩展该研究。这些测量通过我们当地患者压缩乳房厚度分布进行加权,以提供简单的散射因子。提供空气比释动能分布,证明了在存在患者身体衰减的情况下,DBT 系统周围散射辐射的分布。根据当地患者工作量,对每位患者进行 4 视图筛查断层合成检查,在所测量的任何方向上,最高的空气比释动能为 55μGy(距离 1m)。该值可用于保守地将散射辐射视为屏蔽评估的各向同性分布。这项工作与许多最新出版物一致,除了前向散射峰值较小这一明显例外,前向散射峰值比以前的研究小。根据我们当地遇到的乳房厚度分布,在大多数情况下,增加的散射辐射和随之而来的最小屏蔽要求增加很小。对于每周 400 名患者,在 1m 处的空气比释动能为 22mGy,对于距离 2m 的完全占用的公共区域,这需要 18.0mm 厚的石膏板,比以前的数据建议增加 3.6mm。尽管增加幅度很小,但屏蔽材料的标准制造厚度可能不会自动被认为是足够的,因此需要仔细考虑散射辐射。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验