Suppr超能文献

临床数字乳腺断层合成设备周围的散射辐射强度及其对辐射防护考量的影响。

Scatter radiation intensities around a clinical digital breast tomosynthesis unit and the impact on radiation shielding considerations.

作者信息

Yang Kai, Li Xinhua, Liu Bob

机构信息

Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, 55 Fruit Street, Boston, Massachusetts 02114.

出版信息

Med Phys. 2016 Mar;43(3):1096-110. doi: 10.1118/1.4940352.

Abstract

PURPOSE

To measure the scattered radiation intensity around a clinical digital breast tomosynthesis (DBT) unit and to provide updated data for radiation shielding design for DBT systems with tungsten-anode x-ray tubes.

METHODS

The continuous distribution of scattered x-rays from a clinical DBT system (Hologic Selenia Dimensions) was measured within an angular range of 0°-180° using a linear-array x-ray detector (X-Scan 0.8f3-512, Detection Technology, Inc., Finland), which was calibrated for the x-ray spectrum range of the DBT unit. The effects of x-ray field size, phantom size, and x-ray kVp/filter combination were investigated. Following a previously developed methodology by Simpkin, scatter fraction was determined for the DBT system as a function of angle around the phantom center. Detailed calculations of the scatter intensity from a DBT system were demonstrated using the measured scatter fraction data.

RESULTS

For the 30 and 35 kVp acquisition, the scatter-to-primary-ratio and scatter fraction data closely matched with data previously measured by Simpkin. However, the measured data from this study demonstrated the nonisotropic distribution of the scattered radiation around a DBT system, with two strong peaks around 25° and 160°. The majority scatter radiation (>70%) originated from the imaging detector assembly, instead of the phantom. With a workload from a previous survey performed at MGH, the scatter air kerma at 1 m from the phantom center for wall/door is 1.76 × 10(-2) mGy patient(-1), for floor is 1.64 × 10(-1) mGy patient(-1), and for ceiling is 3.66 × 10(-2) mGy patient(-1).

CONCLUSIONS

Comparing to previously measured data for mammographic systems, the scatter air kerma from Holgoic DBT is at least two times higher. The main reasons include the harder primary beam with higher workload (measured with total mAs/week), added tomosynthesis acquisition, and strong small angle forward scattering. Due to the highly conservative initial assumptions, the shielding recommendation from NCRP Report 147 is still sufficient for the Hologic DBT system given the workload from a previous survey at MGH. With the data provided from this study, accurate shielding calculation can be performed for Hologic DBT systems with specific workload and barrier distance.

摘要

目的

测量临床数字乳腺断层合成(DBT)设备周围的散射辐射强度,并为采用钨靶X射线管的DBT系统的辐射屏蔽设计提供最新数据。

方法

使用线性阵列X射线探测器(X-Scan 0.8f3-512,芬兰Detection Technology公司)在0°至180°的角度范围内测量临床DBT系统(Hologic Selenia Dimensions)散射X射线的连续分布,该探测器针对DBT设备的X射线光谱范围进行了校准。研究了X射线野大小、体模大小和X射线管电压/滤过组合的影响。按照Simpkin先前开发的方法,确定DBT系统的散射分数作为围绕体模中心的角度的函数。使用测量的散射分数数据展示了DBT系统散射强度的详细计算。

结果

对于30和35 kVp采集,散射与原发射线比和散射分数数据与Simpkin先前测量的数据紧密匹配。然而,本研究的测量数据表明DBT系统周围散射辐射的非各向同性分布,在25°和160°左右有两个强峰。大部分散射辐射(>70%)源自成像探测器组件,而非体模。根据MGH先前进行的一项调查的工作量,距体模中心1 m处,墙壁/门处的散射空气比释动能为1.76×10⁻² mGy/患者,地板处为1.64×10⁻¹ mGy/患者,天花板处为3.66×10⁻² mGy/患者。

结论

与先前测量的乳腺摄影系统数据相比,Hologic DBT的散射空气比释动能至少高出两倍。主要原因包括具有更高工作量(以每周总mAs测量)的更硬原发射线、增加的断层合成采集以及强烈的小角度前向散射。由于最初的假设高度保守,鉴于MGH先前一项调查的工作量,NCRP报告147中的屏蔽建议对于Hologic DBT系统仍然足够。利用本研究提供的数据,可以针对具有特定工作量和屏蔽距离的Hologic DBT系统进行精确的屏蔽计算。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验