Abstract

INTRODUCTION

Mammographic imaging can cause considerable stress and anxiety for some patients and may require someone to remain in the room during the procedure to provide both physical and emotional support. As such, these comforters and carers (C&C) are exposed to ionising radiation. Limited evidence is available stating the radiation dose received during a digital breast tomosynthesis (DBT) examination. This research aims to determine the optimal standing position for a C&C in the mammography room during a DBT mammogram that results in the lowest radiation dose, whilst providing high-quality imaging, care and comfort to the patient.

METHODS

A scatter detector was used to measure the dose at different standing positions of the carer relative to the patient during an examination. A polymethyl methacrylate (PMMA) phantom was also used to model the patient's breast and torso for further scatter dose measurements.

RESULTS

The median air kerma for craniocaudal views posterior to the patient is 0.75 μGy compared with 10.1 μGy to either side. The median air kerma for mediolateral oblique views for posterolateral position is 0.41 μGy compared with 2.6 μGy anterolateral. No significant effect from breast density is noted from the dataset.

CONCLUSION

The optimal position for the C&C to stand is directly behind the patient in the craniocaudal position, and as far as possible posterolateral to the breast being imaged in the mediolateral oblique position. These two positions will result in the least radiation dose to the C&C.

摘要

简介

乳腺 X 光成像可能会给一些患者带来相当大的压力和焦虑,并且可能需要有人在检查过程中留在房间里,为患者提供身体和情感上的支持。因此,这些安慰者和护理人员(C&C)会暴露在电离辐射下。目前仅有有限的证据表明,在数字乳腺断层合成(DBT)检查期间,C&C 所接受的辐射剂量。本研究旨在确定在 DBT 乳房 X 光检查期间,C&C 在乳房 X 光室中的最佳站立位置,以获得最低的辐射剂量,同时为患者提供高质量的成像、护理和舒适感。

方法

使用散射探测器测量在检查过程中,护理人员相对于患者处于不同站立位置时的剂量。还使用聚甲基丙烯酸甲酯(PMMA)体模对患者的乳房和躯干进行建模,以进一步进行散射剂量测量。

结果

与患者两侧相比,患者后方的头尾位的空气比释动能中位数为 0.75μGy,而侧斜位的后外侧位为 0.41μGy,前外侧位为 2.6μGy。从数据集未注意到乳房密度的显著影响。

结论

C&C 站立的最佳位置是直接在患者后方的头尾位,并且在侧斜位中尽可能靠近要成像的乳房的后外侧位。这两个位置将使 C&C 接受的辐射剂量最小。