全视野数字化乳腺摄影设备周围的散射辐射强度。

Scatter radiation intensities around full-field digital mammography units.

机构信息

BreastCheck, The National Cancer Screening Service, Galway, Ireland.

出版信息

Br J Radiol. 2013 Jan;86(1021):20120130. doi: 10.1259/bjr.20120130.

Abstract

The aim of this study was to investigate the scatter radiation intensity around digital mammography systems and apply these data to standard shielding calculations to reveal whether shielding design of existing breast screening rooms is adequate for the use of digital mammography systems. Three digital mammography systems from GE Healthcare, Hologic and Philips were employed in the study. A breast-equivalent phantom was imaged under clinical workload conditions and scatter radiation intensities around the digital mammography systems were measured for a range of angles in three planes using an ionisation chamber. The results were compared with those from previous studies of film-screen systems. It may be deduced from the results that scattering in the backward direction is significant for all three systems, while scattering in the forward direction can be significant for some planes around the GE and Hologic systems. Measurements at typical clinical settings on each system revealed the Philips system to have markedly lower scatter radiation intensities than the other systems. Substituting the measured scattered radiation intensity into shielding calculations yielded barrier requirements similar to those already in place at the screening centres operating these systems. Current radiation protection requirements based on film-screen technology remain sufficient when applied to rooms with digital mammography installations and no alteration is required to the structural shielding.

摘要

本研究旨在探讨数字化乳腺摄影系统周围的散射辐射强度,并将这些数据应用于标准屏蔽计算,以揭示现有的乳腺筛查室的屏蔽设计是否足以满足数字化乳腺摄影系统的使用需求。本研究使用了来自通用电气医疗、豪洛捷和飞利浦的三种数字化乳腺摄影系统。在临床工作量条件下对乳腺等效体模进行成像,并使用电离室在三个平面的多个角度测量数字化乳腺摄影系统周围的散射辐射强度。结果与之前的胶片-屏系统研究结果进行了比较。从结果中可以推断,所有三种系统的向后散射都很显著,而在某些平面上,GE 和豪洛捷系统的向前散射可能很显著。在每个系统的典型临床设置下进行的测量表明,飞利浦系统的散射辐射强度明显低于其他系统。将测量到的散射辐射强度代入屏蔽计算,得出的屏蔽要求与运行这些系统的筛查中心现有的要求相似。当应用于安装了数字化乳腺摄影设备的房间时,基于胶片-屏技术的现行辐射防护要求仍然足够,不需要对结构屏蔽进行任何更改。

相似文献

本文引用的文献

2
Comprehensive dose survey of breast screening in Ireland.爱尔兰乳房筛查综合剂量调查。
Radiat Prot Dosimetry. 2011 Apr;145(1):52-60. doi: 10.1093/rpd/ncq375. Epub 2010 Nov 18.
5
Scatter radiation intensities about mammography units.乳腺摄影设备周围的散射辐射强度。
Health Phys. 1996 Feb;70(2):238-44. doi: 10.1097/00004032-199602000-00012.
6
Scatter radiation in mammography.乳腺摄影中的散射辐射。
Australas Phys Eng Sci Med. 1993 Mar;16(1):33-6.
8
Transmission data for shielding diagnostic x-ray facilities.屏蔽诊断X射线设施的传输数据。
Health Phys. 1995 May;68(5):704-9. doi: 10.1097/00004032-199505000-00011.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验