Spence P A, Peniston C M, David T E, Mihic N, Jabr A K, Narini P, Salerno T A
Ann Thorac Surg. 1986 Apr;41(4):363-71. doi: 10.1016/s0003-4975(10)62688-4.
The relatively high morbidity and mortality of mitral valve replacement (MVR) appears to be related to an impairment in left ventricular (LV) function. This two-part investigation was designed to assess the effect of MVR on global LV function in an isolated heart preparation and to evaluate whether a mitral prosthesis with a flexible annulus would be of benefit. In Part I (14 pigs), the effects of each step in MVR were studied. Division of the chordae tendineae caused a severe deterioration in LV function (systolic pressure, 180 +/- 13 versus 120 +/- 10 mm Hg; p less than 0.05; developed pressure, 167 +/- 13 versus 108 +/- 11 mm Hg; p less than 0.05; first derivative of LV pressure [dP/dt], 2,630 +/- 300 versus 1,610 +/- 180 mm Hg/sec; p less than 0.05; balloon volume, 30 ml). Fixation of the mitral annulus prior to division of the chordae tendineae resulted in a small decrease (not significant) in LV function but had no effect after the chordae tendineae were divided. In Part II (10 pigs), two mitral annular prostheses were studied: a standard rigid prosthesis and a prosthesis of identical size but with a flexible annulus. LV function was better with the flexible than the rigid prosthesis (systolic pressure, 118 +/- 10 versus 89 +/- 5% control, p less than 0.02; developed pressure, 120 +/- 11 versus 87 +/- 5% control; p less than 0.02; dP/dt, 119 +/- 10 versus 85 +/- 4% control; p less than 0.02; balloon volume, 30 ml).(ABSTRACT TRUNCATED AT 250 WORDS)
二尖瓣置换术(MVR)相对较高的发病率和死亡率似乎与左心室(LV)功能受损有关。这项分为两部分的研究旨在评估在离体心脏标本中MVR对整体左心室功能的影响,并评估具有柔性瓣环的二尖瓣假体是否有益。在第一部分(14头猪)中,研究了MVR每个步骤的影响。腱索离断导致左心室功能严重恶化(收缩压,180±13对120±10 mmHg;p<0.05;舒张末压,167±13对108±11 mmHg;p<0.05;左心室压力一阶导数[dP/dt],2630±300对1610±180 mmHg/秒;p<0.05;球囊容积,30 ml)。在腱索离断前固定二尖瓣瓣环导致左心室功能有小幅下降(不显著),但在腱索离断后没有影响。在第二部分(10头猪)中,研究了两种二尖瓣瓣环假体:一种标准刚性假体和一种尺寸相同但瓣环柔性的假体。柔性假体的左心室功能优于刚性假体(收缩压,118±10对89±5%对照,p<0.02;舒张末压,120±11对87±5%对照;p<0.02;dP/dt,119±10对85±4%对照;p<0.02;球囊容积,30 ml)。(摘要截断于250字)