Blood gas, electrolyte, glucose, and lactate level measurement have an immediate and critical impact on patient care. We evaluated the performance of i-SmartCare 10 (i-SENS Inc., Seoul, Korea) and conducted a method comparison study of five point-of-care (POC) analyzers with i-SmartCare 10 as the comparator, according to the CLSI guidelines. Ten analytes (pH, pCO, pO, Na, K, Cl, iCa, glucose, lactate, and Hct) were tested on six analyzers: i-SmartCare 10, ABL90 FLEX PLUS (Radiometer Medical ApS, Copenhagen, Denmark), i-Stat (Abbott Point of Care Inc., Princeton, NJ, USA), RapidLab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA), Stat Profile pHOx Ultra (Nova Biomedical, Waltham, MA, USA), and Gem Premier 5000 (Instrumentation Laboratory, Bedford, MA, USA). The total imprecision and linearity (r>0.99) were excellent, except for a few analytes that narrowly escaped the preset criteria. Interference was noted for Na in the presence of a high K level and for iCa in the presence of high K and Mg levels. Forty of 48 items demonstrated either a proportional or systematic difference in regression analysis; the relative mean difference (%) of 14/48 items escaped the allowable total error in the difference plot analysis. i-SmartCare 10 shows acceptable performance, and using a single POC blood gas analyzer is recommended for monitoring.
血气、电解质、血糖和乳酸水平的测量对患者的治疗具有直接而关键的影响。我们评估了 i-SmartCare 10(i-SENS Inc.,韩国首尔)的性能,并根据 CLSI 指南对 5 种即时检验(POC)分析仪与 i-SmartCare 10 进行了方法比较研究。在 6 种分析仪上测试了 10 种分析物(pH、pCO、pO、Na、K、Cl、iCa、葡萄糖、乳酸和 Hct):i-SmartCare 10、ABL90 FLEX PLUS(Radiometer Medical ApS,哥本哈根,丹麦)、i-Stat(雅培 POCT 公司,新泽西州普林斯顿)、RapidLab 1265(西门子医疗诊断公司,纽约州塔里敦)、Stat Profile pHOx Ultra(美国新百伦公司,沃尔瑟姆)和 Gem Premier 5000(美国仪器实验室公司,贝德福德)。除了少数分析物勉强符合预设标准外,总不精密度和线性(r>0.99)都非常好。在高钾水平存在的情况下,Na 存在干扰,在高钾和镁水平存在的情况下,iCa 存在干扰。在 48 项中有 40 项显示出回归分析中的比例或系统差异;在差异图分析中,14/48 项的相对平均差异(%)超出了可允许的总误差。i-SmartCare 10 的性能可接受,建议使用单一的即时血气分析仪进行监测。