企业即时检测分析仪与诺瓦生物医学重症监护快速分析仪用于测定犬马血液中动脉血pH值、血气和电解质值的比较。

Comparison of Enterprise Point-of-Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood.

作者信息

Elmeshreghi Taher N, Grubb Tamara L, Greene Stephen A, Ragle Claude A, Wardrop Jane A

机构信息

Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Washington State University, Pullman, WA, USA.

出版信息

Vet Clin Pathol. 2018 Sep;47(3):415-424. doi: 10.1111/vcp.12635. Epub 2018 Jul 9.

Abstract

BACKGROUND

Point-of-care analyzers can provide a rapid turnaround time for critical blood test results. Agreement between the Enterprise Point-of-Care (EPOC) and bench-top laboratory analyzers is important to determine the clinical reliability of the EPOC.

OBJECTIVES

The aim of the study was (1) to evaluate the precision (repeatability) of blood gas values measured by the EPOC and (2) to determine the level of agreement between the EPOC and Nova Critical Care Express (Nova CCX) for the assessment of arterial pH, blood gases, and electrolyte variables in canine and equine blood.

METHODS

Arterial blood samples from dogs were analyzed on the EPOC and Nova CCX analyzers to determine precision and agreement of pH, PaCO , PaO , and HCT. The same analytes plus Na , K , and Cl were analyzed for agreement using equine blood. Statistical analyses included assessment of precision using the coefficient of variation (CV%), and agreement using the Deming regression, Pearson correlation, and Bland-Altman plots.

RESULTS

Both analyzers provided precise results of pH, PaCO , PaO and HCT, meeting CV% quality requirement values. In both species, Deming regression results were acceptable and correlation values were above 0.93 for arterial pH and blood gases, but lower for sodium and chloride. Bland-Altman plots demonstrated varying degrees of bias, but good agreement between the 2 analyzers was seen when arterial blood gases and electrolytes were measured, except for PaCO and Cl CONCLUSION: The EPOC analyzer provides consistent, reliable results for canine arterial blood gas values and for equine arterial blood gas and electrolyte values. Cl results could be acceptable with the application of a correction factor, but the PaCO results were more variable.

摘要

背景

即时检验分析仪能够快速给出关键血液检测结果。企业即时检验(EPOC)分析仪与台式实验室分析仪之间的一致性对于确定EPOC的临床可靠性至关重要。

目的

本研究的目的是(1)评估EPOC测量血气值的精密度(重复性),以及(2)确定EPOC与诺瓦重症监护快速分析仪(Nova CCX)在评估犬和马血液中的动脉血pH值、血气和电解质变量方面的一致性水平。

方法

使用EPOC和Nova CCX分析仪对犬的动脉血样本进行分析,以确定pH值、二氧化碳分压(PaCO₂)、氧分压(PaO₂)和血细胞比容(HCT)的精密度和一致性。使用马血对相同的分析物以及钠(Na⁺)、钾(K⁺)和氯(Cl⁻)进行一致性分析。统计分析包括使用变异系数(CV%)评估精密度,以及使用戴明回归、皮尔逊相关和布兰德-奥特曼图评估一致性。

结果

两种分析仪都提供了精确的pH值、PaCO₂、PaO₂和HCT结果,符合CV%质量要求值。在两个物种中,戴明回归结果均可接受,动脉血pH值和血气的相关值高于0.93,但钠和氯的相关值较低。布兰德-奥特曼图显示出不同程度的偏差,但在测量动脉血气和电解质时,两种分析仪之间有良好的一致性,除了PaCO₂和Cl⁻。

结论

EPOC分析仪为犬动脉血气值以及马动脉血气和电解质值提供了一致、可靠的结果。应用校正因子时,Cl⁻结果可能可以接受,但PaCO₂结果的变异性更大。