Inje University College of Medicine, Busan, Korea.
Department of Clinical Laboratory Science, Semyung University, Jecheon, Korea.
Ann Lab Med. 2022 Jul 1;42(4):478-481. doi: 10.3343/alm.2022.42.4.478.
Group B streptococcus (GBS) is an important pathogen causing neonatal early-onset disease. We evaluated the diagnostic performance of BD Max GBS assay (Becton Dickinson, Franklin Lakes, NJ, USA) without enrichment (direct BDM) for detecting GBS using vaginal and rectal specimens in comparison with culture. In total, 716 specimens collected from 358 pregnant women between June 2018 and May 2020 were included in this study. Bacterial culture was performed using ChromID Strep B agar (bioMérieux, Marcy-l'Étoile, France), and species identification results were confirmed using the VITEK-MS system (bioMérieux). The sensitivity of direct BDM for vaginal and rectal specimens was 75.0% and 100%, respectively. Thirteen specimens showed discrepant results: 10 false-negative results in the vaginal specimens and three false-positive results in the rectal specimens. The overall agreement between direct BDM and culture was 98.9% (354/358). The final sensitivity and specificity of direct BDM were 98.5% and 99.0%, respectively. Discrepant results-one false-negative and three false-positives-were obtained for four specimens. Direct BDM shows a good diagnostic performance and will be useful for GBS screening within a few hours.
B 群链球菌(GBS)是一种引起新生儿早发性疾病的重要病原体。我们评估了 BD Max GBS 检测试剂盒(Becton Dickinson,Franklin Lakes,NJ,USA)在未经富集的情况下(直接 BDM)检测阴道和直肠标本中 GBS 的诊断性能,并与培养进行了比较。本研究共纳入了 2018 年 6 月至 2020 年 5 月期间 358 名孕妇的 716 份标本。细菌培养使用 ChromID Strep B 琼脂(bioMérieux,Marcy-l'Étoile,France)进行,物种鉴定结果使用 VITEK-MS 系统(bioMérieux)确认。直接 BDM 对阴道和直肠标本的敏感性分别为 75.0%和 100.0%。13 份标本的结果不一致:阴道标本中有 10 份假阴性,直肠标本中有 3 份假阳性。直接 BDM 与培养的总符合率为 98.9%(354/358)。直接 BDM 的最终敏感性和特异性分别为 98.5%和 99.0%。4 份标本出现不一致的结果-1 份假阴性和 3 份假阳性。直接 BDM 显示出良好的诊断性能,将在数小时内有助于 GBS 筛查。