Hoffmann Chelsey M, Coy Debra S, Moeschler Susan M, Pingree Matthew J, Mauck William D
Department of Anesthesiology and Perioperative Medicine, Division of Pain Medicine, Mayo Clinic, Rochester, MN, USA.
Department of Anesthesiology and Perioperative Medicine, Division of Pain Medicine, Mayo Clinic, Rochester, MN, USA.
Neuromodulation. 2023 Apr;26(3):690-693. doi: 10.1016/j.neurom.2022.01.018. Epub 2022 Feb 26.
The primary objective of this study was to determine the current rate of lead fracture during temporary percutaneous peripheral nerve stimulator (PNS) lead removal at the Mayo Clinic Rochester Division of Pain Medicine.
A retrospective review of electronic medical records was performed for patients implanted with a temporary percutaneous PNS device between January 1, 2018, and December 31, 2020. Patients were included if they underwent temporary percutaneous PNS system implant, with planned lead removal at 60 days. Data collection included date of implant, diagnosis, peripheral nerve target(s), number of leads, and lead tip status at the time of removal (intact vs fractured).
Fifty patients underwent a total of 80 temporary percutaneous PNS leads placed during the time frame analyzed. Of the 80 temporary percutaneous PNS leads implanted, there were five lead fractures at the time of intentional lead removal.
This retrospective review of 50 patients with 80 temporary percutaneous PNS leads implanted for chronic peripheral neuropathic pain resulted in a 6.25% rate of retained lead fragment at the time of lead removal by the provider at the end of 60-day treatment. This fracture and retention rate is consistent with previous published retrospective data on PNS, which has shown a 3% to 21% lead fracture rate during intentional lead removal.
本研究的主要目的是确定梅奥诊所罗切斯特疼痛医学科在临时经皮周围神经刺激器(PNS)导线移除过程中当前的导线断裂率。
对2018年1月1日至2020年12月31日期间植入临时经皮PNS设备的患者的电子病历进行回顾性分析。纳入标准为接受临时经皮PNS系统植入且计划在60天时移除导线的患者。数据收集包括植入日期、诊断、周围神经靶点、导线数量以及移除时的导线尖端状态(完整或断裂)。
在分析的时间段内,50例患者共植入了80根临时经皮PNS导线。在这80根植入的临时经皮PNS导线中,有5根在有意移除导线时发生了断裂。
对50例植入80根临时经皮PNS导线用于慢性周围神经病理性疼痛的患者进行的这项回顾性研究表明,在60天治疗结束时由提供者移除导线时,导线碎片残留率为6.25%。这一断裂和残留率与先前发表的关于PNS的回顾性数据一致,该数据显示在有意移除导线期间,导线断裂率为3%至21%。