Suppr超能文献

带有嵌入式传感器手柄和新型混合式超声乳化头的白内障手术的疗效和安全性。

The efficacy and safety profile of cataract procedures performed with a sensor-embedded handpiece and a new hybrid phaco tip.

机构信息

Department of Ophthalmology, Izzet Baysal State Hospital, Bolu, Turkey.

Department of Ophthalmology, Batigoz Eye Health Center, Izmir, Turkey.

出版信息

Eur J Ophthalmol. 2022 Nov;32(6):3438-3443. doi: 10.1177/11206721221084764. Epub 2022 Mar 1.

Abstract

OBJECTIVE

To evaluate the efficacy and safety of the new Intrepid Hybrid tip and Active Sentry handpiece in comparison to the Intrepid Balanced tip and Centurion Ozil handpiece.

METHODS

One hundred sixty-eight eyes were included in the study and randomized into Group 1 (Active Sentry handpiece + Intrepid Hybrid tip, n = 86) and Group 2 (Centurion Ozil handpiece + Intrepid Balanced tip, n = 82). Data were collected from the phaco device, including total U/S time, cumulative dissipated energy (CDE), the total number of active surge mitigation actuations (ASM actuations), torsional amplitude, torsional amplitude on time, estimated fluid of aspirated, and aspiration time. Endothelial cell count (ECC) and central corneal thickness (CCT) were assessed. A correlation analysis was performed between ASM actuations, phaco metrics, and patient characteristics in Group 1.

RESULTS

The CDE, torsional amplitude measured in Group 1 (8.8 ± 3.9, 51.2 ± 13.3, respectively) were significantly lower than those of Group 2 (10.4 ± 4.2, 65.2 ± 9.3, respectively). While no complications were observed in group 1, there were posterior capsule rupture in 2 eyes and iris damage in 1 eye in group 2. The mean ECC and CCT were similar at 1 month. The ASM actuations were more engaged in the eyes with pseudoexfoliation and small pupils.

CONCLUSION

The combination of the new Intrepid Hybrid tip and Active Sentry handpiece enables safer cataract procedures without compromising efficacy due to its surge prevention mechanisms and special tip design. It could be a good option for residents in training.

摘要

目的

评估新型 Intrepid Hybrid 尖端和 Active Sentry 手柄与 Intrepid Balanced 尖端和 Centurion Ozil 手柄相比的疗效和安全性。

方法

这项研究纳入了 168 只眼,并将其随机分为第 1 组(Active Sentry 手柄+Intrepid Hybrid 尖端,n=86)和第 2 组(Centurion Ozil 手柄+Intrepid Balanced 尖端,n=82)。从超声乳化仪收集数据,包括总超声时间、累积耗散能量(CDE)、主动声速降低激活次数(ASM 激活次数)、扭转幅度、扭转幅度持续时间、预估抽吸的液体量和抽吸时间。评估内皮细胞计数(ECC)和中央角膜厚度(CCT)。对第 1 组中 ASM 激活次数、超声乳化仪参数与患者特征进行了相关性分析。

结果

第 1 组的 CDE、扭转幅度分别为 8.8±3.9、51.2±13.3,显著低于第 2 组的 10.4±4.2、65.2±9.3。第 1 组无并发症发生,第 2 组有 2 只眼发生后囊破裂和 1 只眼发生虹膜损伤。术后 1 个月,平均 ECC 和 CCT 相似。在有假性剥脱和瞳孔较小的眼中,ASM 激活次数更多。

结论

新型 Intrepid Hybrid 尖端和 Active Sentry 手柄的结合通过其防声速骤升机制和特殊尖端设计,实现了更安全的白内障手术,同时不影响疗效,可为培训中的住院医师提供一个不错的选择。