[数字影像增强器X线摄影术:成像质量及在胃肠道区域的初步应用]

[Digital image intensifier radiography: imaging qualities and initial use in the gastrointestinal area].

作者信息

Busch H P, Strauss L G, Prager P, Georgi M

出版信息

Digitale Bilddiagn. 1986 Jun;6(2):49-55.

PMID:3524964
Abstract

We determined the quality of digital radiographs delivered by a large image intensifier (diameter 54 cm) and by a conventional fluoroscopic unit with a 33 cm image intensifier, and compared them to that of 100 mm spot films obtained from the same units. A special image processor was available allowing 25 images/s using a 512/512 matrix or 7.5 images/s with a 1024 X 1024 matrix. The spatial resolution of digital radiographs was well below that of the spot films, while no difference was found in contrast resolution. Images of colon specimens and gastrointestinal studies demonstrate that digital radiography does not imply a loss in diagnostic information inspite of the lower spatial resolution. Gastrointestinal studies using digital radiographs will reduce the amount of irradiation exposure of the patient and could represent an additional use for an existing equipment for DSA.

摘要

我们测定了由大型影像增强器(直径54厘米)和配备33厘米影像增强器的传统荧光透视装置所提供的数字X线片的质量,并将它们与从相同装置获得的100毫米点片的质量进行比较。有一台特殊的影像处理器,使用512/512矩阵时每秒可处理25幅图像,使用1024×1024矩阵时每秒可处理7.5幅图像。数字X线片的空间分辨率远低于点片,而在对比分辨率方面未发现差异。结肠标本图像和胃肠道研究表明,尽管空间分辨率较低,但数字X线摄影并不意味着诊断信息的丢失。使用数字X线片进行胃肠道研究将减少患者的辐射暴露量,并且可以成为现有数字减影血管造影(DSA)设备的一种额外用途。