测量慢性运动性骨筋膜室综合征的骨筋膜室内压力:对市售设备及其各自准确性的挑战。

Measuring intracompartmental pressures for the chronic exertional compartment syndrome: Challenging commercially available devices and their respective accuracy.

作者信息

Vogels S, Ritchie E D, Bakker E W P, Vogels M A J M, Zimmermann W O, Verhofstad M H J, Hoencamp R

机构信息

Department of Surgery, Alrijne Hospital, Leiderdorp, the Netherlands; Trauma Research Unit, Department of Trauma Surgery, Erasmus Medical Center, Rotterdam, the Netherlands.

Department of Surgery, Alrijne Hospital, Leiderdorp, the Netherlands.

出版信息

J Biomech. 2022 Apr;135:111026. doi: 10.1016/j.jbiomech.2022.111026. Epub 2022 Feb 28.

Abstract

The aim of this study was to compare commercially available manometers and needles used for intracompartmental pressure measurements for accuracy. An experimental compartment simulation model was developed in order to compare four different terminal devices (Compass manometer, Stryker device, Meritrans transducer, and arterial line) and 22 types of needles. First, all possible device/needle combinations were introduced in rubber ports at the bottom of a water column. The water column was gradually drained three times for each device/needle combination and pressures were recorded by two researchers. This procedure was repeated after placement of a sample of homogenous porcine gluteal muscle tissue. When measuring the fluid column only, all Intraclass Correlation Coefficients were found to be ≥0.980, indicating good resemblance to the reference pressure. After addition of the muscle tissue to the experimental set up, accuracy of the Stryker device and arterial line remained the same. The accuracy of the Compass manometer and Meritrans transducer became less. Excellent reliability was found for 7 out of 22 needles when combined with the Stryker device and for 2 out of 22 needles when combined with the arterial line. The four terminal devices tested were all accurate when measuring pressure in a water column, whereas only the Stryker device and arterial line accurately represented the water column pressure in a porcine gluteal muscle model. Future research will have to verify whether these results can be repeated in human cadaver models and in vivo.

摘要

本研究的目的是比较用于测量骨筋膜室内压力的市售压力计和针头的准确性。为了比较四种不同的终端设备(Compass压力计、史赛克设备、Meritrans传感器和动脉导管)以及22种针头,开发了一种实验性的骨筋膜室模拟模型。首先,将所有可能的设备/针头组合引入水柱底部的橡胶端口。对于每种设备/针头组合,将水柱逐渐排空三次,并由两名研究人员记录压力。在放置均匀的猪臀肌组织样本后,重复此过程。仅测量液柱时,所有组内相关系数均≥0.980,表明与参考压力具有良好的相似性。在实验装置中加入肌肉组织后,史赛克设备和动脉导管的准确性保持不变。Compass压力计和Meritrans传感器的准确性降低。与史赛克设备组合时,22种针头中有7种具有出色的可靠性;与动脉导管组合时,22种针头中有2种具有出色的可靠性。测试的四种终端设备在测量水柱压力时均准确,而在猪臀肌模型中,只有史赛克设备和动脉导管能准确反映水柱压力。未来的研究将必须验证这些结果是否能在人体尸体模型和体内重复。