Lee Jae Jung, Lee Ji Eun, Sadda Srinivas R, Park Sung Who, Byon Iksoo
Department of Ophthalmology, Pusan National University School of Medicine, Yangsan, South Korea.
Biomedical Research Institute, Pusan National University Hospital, Busan, South Korea.
Sci Rep. 2022 Mar 18;12(1):4692. doi: 10.1038/s41598-022-08781-1.
We evaluated the impact of signal strength (SS) on quantitative measurements from optical coherence tomography (OCTA). Twenty healthy-volunteers were included. A neutral density filter (NDF) was attached to spectral-domain OCTA (SD-OCTA) and swept-source OCTA (SS-OCTA). All subjects were imaged with both devices three times using three different conditions: no filter, NDF0.3, and NDF0.6. For SD-OCTA, SS decreased from 10.0 to 8.2 and 4.0 with the NDF0.3 and 0.6, respectively. The vessel density (VD) and vessel length density (VLD) of the superficial capillary plexus (SCP) decreased when SS decreased from 10 to 8.2, but no further decrease when SS changed from 8.2 to 4.0. The flow metrics of the deep capillary plexus (DCP) did not change. For SS-OCTA, SS decreased from 10 to 9.5 and 7.2. The VD and VLD of the SCP and DCP decreased when SS decreased, except for the VD of the DCP when SS changed from 10 to 9.5. The choriocapillaris flow deficits significantly increased along with the decrease in SS. Quantitative flow parameters were significantly affected by a small change in SS and were most conspicuous in the SCP and choriocapillaris. These finding highlight the importance of high and consistent SS in quantitative OCTA studies.
我们评估了信号强度(SS)对光学相干断层扫描血管造影(OCTA)定量测量的影响。纳入了20名健康志愿者。将中性密度滤光片(NDF)附着于光谱域OCTA(SD - OCTA)和扫频源OCTA(SS - OCTA)。所有受试者使用这两种设备在三种不同条件下各成像三次:无滤光片、NDF0.3和NDF0.6。对于SD - OCTA,使用NDF0.3和NDF0.6时,SS分别从10.0降至8.2和4.0。当SS从10降至8.2时,浅表毛细血管丛(SCP)的血管密度(VD)和血管长度密度(VLD)降低,但当SS从8.2变为4.0时没有进一步降低。深层毛细血管丛(DCP)的血流指标没有变化。对于SS - OCTA,SS从10降至9.5和7.2。当SS降低时,SCP和DCP的VD和VLD降低,但当SS从10变为9.5时,DCP的VD除外。脉络膜毛细血管血流缺损随SS降低而显著增加。SS的微小变化对定量血流参数有显著影响,在SCP和脉络膜毛细血管中最为明显。这些发现凸显了在定量OCTA研究中保持高且一致的SS的重要性。