Audiology, Universitätsklinikum Erlangen, Erlangen, Germany.
Otorhinolaryngology - Head and Neck Surgery, UKSH Kiel, Kiel, Germany.
Audiol Neurootol. 2022;27(5):347-355. doi: 10.1159/000523784. Epub 2022 Mar 18.
Transimpedance measurements from cochlear implant electrodes have the potential to identify anomalous electrode array placement, such as tip fold-over (TFO) or fold-back, basal electrode kinking, or buckling. Analysing transimpedance may thus replace intraoperative or post-operative radiological imaging to detect any potential misplacements. A transimpedance algorithm was previously developed to detect deviations from a normal electrode position with the aim of intraoperatively detecting TFO. The algorithm had been calibrated on 35 forced, tip folded electrode arrays in six temporal bones to determine the threshold criterion required to achieve a sensitivity of 100%. Our primary objective here was to estimate the specificity of this TFO algorithm in patients, in a prospective study, for a series of electrode arrays shown to be normally inserted by post-operative imaging.
Intracochlear voltages were intraoperatively recorded for 157 ears, using Cochlear's Custom Sound™ EP 5 electrophysiological software (Cochlear Ltd., Sydney, NSW, Australia), for both Nucleus® CI512 and CI532 electrode arrays. The algorithm analysed the recorded 22 × 22 transimpedance matrix (TIM) and results were displayed as a heatmap intraoperatively, only visible to the technician in the operating theatre. After all clinical data were collected, the algorithm was evaluated on the bench. The algorithm measures the transimpedance gradients and corresponding phase angles (θ) throughout the TIM and calculates the gradient phase range. If this was greater than the predetermined threshold, the algorithm classified the electrode array insertion as having a TFO.
Five ears had no intraoperative TIM and four anomalous matrices were identified from heatmaps and removed from the specificity analysis. Using the 148 remaining data sets (n = 103 CI532 and n = 45 CI512), the algorithm had an average specificity of 98.6% (95.80%-99.75%).
The algorithm was found to be an effective screening tool for the identification of TFOs. Its specificity was within acceptable levels and resulted in a positive predictive value of 76%, with an estimated incidence of fold-over of 4% in perimodiolar arrays. This would mean 3 out of 4 cases flagged as a fold-over would be correctly identified by the algorithm, with the other being a false positive. The measurements were applied easily in theatre allowing it to be used as a routine clinical tool for confirming correct electrode placement.
从人工耳蜗电极的跨阻测量中,有可能识别出异常的电极阵列放置,例如尖端折叠(TFO)或折叠回,基底电极弯曲或卷曲。因此,分析跨阻可能会替代术中或术后的放射影像学成像,以检测任何潜在的错位。之前已经开发了一种跨阻算法来检测偏离正常电极位置的偏差,目的是在术中检测到 TFO。该算法已经在六块颞骨中的 35 个强制尖端折叠电极阵列上进行了校准,以确定实现 100%灵敏度所需的阈值标准。我们的主要目标是在一项前瞻性研究中,在一系列术后影像学显示正常插入的电极阵列中,估计该 TFO 算法在患者中的特异性。
使用 Cochlear 的 Custom Sound™ EP 5 电生理软件(Cochlear Ltd.,悉尼,新南威尔士州,澳大利亚),对 157 只耳朵进行术中记录的内耳蜗电压,用于 Nucleus® CI512 和 CI532 电极阵列。该算法分析记录的 22×22 跨阻矩阵(TIM),并在术中仅显示给手术室技术员的热图上显示结果。在收集完所有临床数据后,在台架上评估该算法。该算法测量 TIM 中的跨阻梯度和相应的相位角(θ),并计算梯度相位范围。如果超过预定的阈值,算法将电极阵列的插入分类为具有 TFO。
五耳术中没有 TIM,从热图中识别出四个异常矩阵并从特异性分析中删除。使用 148 个剩余数据集(n=103 CI532 和 n=45 CI512),该算法的平均特异性为 98.6%(95.80%-99.75%)。
该算法被发现是识别 TFO 的有效筛选工具。其特异性在可接受的水平内,阳性预测值为 76%,预计在偏中心电极阵列中折叠的发生率为 4%。这意味着算法正确识别的 4 例折叠中的 3 例,另 1 例为假阳性。该测量在手术室内轻松应用,使其成为确认电极放置正确的常规临床工具。