Sultan Salahaden R, Alghamdi Amer, Abdeen Rawan, Almutairi Fahad
Department of Diagnostic Radiology, Faculty of Applied Medical Sciences, King Abdulaziz University, Jeddah, Saudi Arabia.
Department of Diagnostic Radiology, King Fahad Hospital, Al-Baha, Saudi Arabia.
Ultrasonography. 2022 Apr;41(2):291-297. doi: 10.14366/usg.21114. Epub 2021 Jul 31.
To date, limited studies have specifically addressed the reliability of ultrasound point shear-wave elastography (pSWE). Therefore, the aim of the present study was to assess the reproducibility of ultrasound pSWE within and between operators using two ultrasound scanners.
iU22 and EPIQ7 ultrasound scanners were used to assess the reliability of pSWE measurements of four inclusions [L I (8 kPa), L II (14 kPa), L III (48 kPa), and L IV (80 kPa)] at a depth of 3.5 cm in an elasticity phantom using a curvilinear 5-1 MHz transducer. The intraoperator, inter-operator, and inter-scanner reproducibility of pSWE was assessed using intraclass correlation coefficients (ICCs). Bland-Altman plots were used to establish bias and limits of agreement (LoA) between measurements. The accuracy of pSWE from manufacturer values was determined using the one-sample t-test.
Intra-operator agreement was excellent, with an ICC >0.90. The bias in measurements for operator A was -0.36±3.13 kPa (LoA, -6.47 to 5.75), and for operator B it was 1.97±6.29 kPa (LoA, -10.25 to 14.21). Inter-operator agreement was excellent, with an ICC of 0.95. The bias in measurements between operators was -0.42±5.00 kPa (LoA, -10.24 to 9.38). The inter-scanner agreement between EPIQ7 and iU22 was excellent, with an ICC of 0.96. The bias in measurements between scanners was 1.74±4.44 kPa (LoA, -6.95 to 10.45). There was significant overestimation for L I (17.75%) and L II (31.14%) and underestimation for L III (-15.28%) and L VI (-98.00%) relative to the manufacturer-reported values.
Phantom ultrasound pSWE was reproducible within and between operators, and between Philips ultrasound scanners; further studies using different ultrasound systems and transducers are required.
迄今为止,专门针对超声点剪切波弹性成像(pSWE)可靠性的研究有限。因此,本研究的目的是使用两台超声扫描仪评估超声pSWE在操作者内部以及操作者之间的可重复性。
使用iU22和EPIQ7超声扫描仪,采用曲线5-1MHz探头,在弹性体模中深度为3.5cm处评估四种内含物[L I(8kPa)、L II(14kPa)、L III(48kPa)和L IV(80kPa)]的pSWE测量的可靠性。使用组内相关系数(ICC)评估pSWE的操作者内、操作者间和扫描仪间的可重复性。使用Bland-Altman图确定测量之间的偏差和一致性界限(LoA)。使用单样本t检验确定pSWE相对于制造商值的准确性。
操作者内一致性极佳,ICC>0.90。操作者A测量的偏差为-0.36±3.13kPa(LoA,-6.47至5.75),操作者B测量的偏差为1.97±6.29kPa(LoA,-10.25至14.21)。操作者间一致性极佳,ICC为0.95。操作者之间测量的偏差为-0.42±5.00kPa(LoA,-10.24至9.38)。EPIQ7和iU22之间的扫描仪间一致性极佳,ICC为0.96。扫描仪之间测量的偏差为1.74±4.44kPa(LoA,-6.95至10.45)。相对于制造商报告的值,L I(17.75%)和L II(31.14%)有显著高估,L III(-15.28%)和L VI(-98.00%)有显著低估。
体模超声pSWE在操作者内部、操作者之间以及飞利浦超声扫描仪之间具有可重复性;需要使用不同超声系统和探头进行进一步研究。