Wang Hanyu, Wang Meng, Wang Hui, Ren Xiang, Wang Huan, Wei Qin, Wu Dan
School of Chemistry and Chemical Engineering, University of Jinan, Jinan, 250022, PR China.
Hand and Foot Surgery Department, Shandong Provincial Hospital affiliated to Shandong First Medical University, Jinan, 250021, PR China.
Biosens Bioelectron. 2022 Jul 1;207:114196. doi: 10.1016/j.bios.2022.114196. Epub 2022 Mar 18.
Generally, the photoanodic photoelectrochemical (PEC) immunoassay method has an outstanding photocurrent and low detection limit, but its poor anti-interference ability in the detection of real samples restricts its performance. The photocathode immunoassay method has an excellent ability to see interference in actual sample detection, but it has its own defect in that the photocurrent is not obvious. Here, a promising new cathodic PEC immunosensing platform is reported, which integrates a photocathode and photoanode. The photoanode and photocathode are WO/Mn:CdS composite modified and reduced graphene oxide (RGO) modified indium tin oxide (ITO) electrodes, respectively. In addition to an excellent PEC response, the immunosensor constructed by the integrating the photoanode and photocathode also has good anti-interference ability in actual sample analysis. The constructed immunosensor achieves accurate detection of NSE with a range from 5.0 pg/mL to 20 ng/mL, and the limit of detection (LOD) is 1.2 pg/mL. The proposed immunoassay method has good stability, selectivity and reproducibility. Moreover, it introduces new ideas for the construction of PEC immunosensors.
一般来说,光阳极光电化学(PEC)免疫分析方法具有出色的光电流和低检测限,但其在实际样品检测中的抗干扰能力较差,限制了其性能。光阴极免疫分析方法在实际样品检测中具有出色的抗干扰能力,但存在光电流不明显的缺陷。在此,报道了一种有前景的新型阴极PEC免疫传感平台,它集成了光阳极和光阴极。光阳极和光阴极分别是WO/Mn:CdS复合材料修饰的和还原氧化石墨烯(RGO)修饰的氧化铟锡(ITO)电极。除了出色的PEC响应外,通过集成光阳极和光阴极构建的免疫传感器在实际样品分析中也具有良好的抗干扰能力。构建的免疫传感器实现了对NSE在5.0 pg/mL至20 ng/mL范围内的准确检测,检测限(LOD)为1.2 pg/mL。所提出的免疫分析方法具有良好的稳定性、选择性和重现性。此外,它为PEC免疫传感器的构建引入了新的思路。