应用未层压的 EBT3 胶片剂量仪定量测量含银纳米粒子海藻酸盐膜的剂量增强效应。

Application of unlaminated EBT3 film dosimeter for quantification of dose enhancement using silver nanoparticle-embedded alginate film.

机构信息

Radiological Physics & Advisory Division, Bhabha Atomic Research Centre, Mumbai-400094, India.

Homi Bhabha National Institute, Anushaktinagar, Mumbai-400094, India.

出版信息

Biomed Phys Eng Express. 2022 Apr 5;8(3). doi: 10.1088/2057-1976/ac60c5.

Abstract

The paper describes the application of unlaminated Gafchromic EBT3 film dosimeter for quantification of dose enhancement using locally synthesized silver nanoparticle-embedded alginate film (AgNPs-Alg film) for nanoparticles-aided radiotherapy.AgNPs-Alg film was synthesized and characterized using standard techniques. The unlaminated Gafchromic EBT3 film was specially customized for dosimetric measurement. The dose enhancements due to AgNPs-Alg film was experimentally determined for ISO wide spectrum x-rays series (average energy ranging from 57-137 keV) and 6 and 10 MV x-rays using laminated and unlaminated Gafchromic EBT3 film. The radiation dose of 1 Gy was delivered to a combination of AgNPs-Alg films and EBT3 film.Ultraviolet-Visible spectroscopy of silver nanoparticles shows a surface plasmon resonance peak at 400 nm. The average particle size of 13 ± 2 nm was measured using Atomic Force Microscopy. For unlaminated film, the dose enhancements of 29%, 23%, 14% and 2% was observed for ISO wide spectrum x-rays having average energy of 57, 79, 104 and 137 keV, respectively. The dose enhancement was negligible for 6 and 10 MV x-rays. In the case of laminated film, no significant dose enhancement was measured for all the x-ray energies.The unlaminated Gafchromic EBT3 film can be a suitable choice for the measurement of dose enhancement. Further, silver nanoparticles can be used during nanoparticle-aided radiotherapy when irradiated at low x-ray energy.

摘要

本文描述了未层压的 Gafchromic EBT3 胶片剂量计在使用本地合成的银纳米粒子嵌入海藻酸盐膜(AgNPs-Alg 膜)进行纳米颗粒辅助放射治疗的剂量增强定量中的应用。AgNPs-Alg 膜是使用标准技术合成和表征的。未层压的 Gafchromic EBT3 薄膜是专门为剂量测量定制的。使用层压和未层压的 Gafchromic EBT3 薄膜,实验确定了 ISO 宽光谱 X 射线系列(平均能量范围为 57-137keV)和 6 和 10MV X 射线中由于 AgNPs-Alg 膜引起的剂量增强。将 1Gy 的辐射剂量输送到 AgNPs-Alg 膜和 EBT3 膜的组合中。银纳米粒子的紫外可见光谱显示出 400nm 的表面等离子体共振峰。使用原子力显微镜测量平均粒径为 13±2nm。对于未层压的薄膜,对于平均能量分别为 57、79、104 和 137keV 的 ISO 宽光谱 X 射线,观察到 29%、23%、14%和 2%的剂量增强。6 和 10MV X 射线的剂量增强可以忽略不计。对于层压薄膜,对于所有 X 射线能量,都没有测量到明显的剂量增强。未层压的 Gafchromic EBT3 薄膜可以是测量剂量增强的合适选择。此外,当在低 X 射线能量下辐照时,银纳米粒子可以在纳米颗粒辅助放射治疗中使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验