Zhou Wenhan, Li Xinyu, Yao Fanglan, Zhang Haozhi, Sun Ke, Chen Fang, Xu Pengcheng, Li Xinxin
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China.
School of Microelectronics, University of Chinese Academy of Sciences, Beijing 100049, China.
Micromachines (Basel). 2022 Mar 16;13(3):445. doi: 10.3390/mi13030445.
Combined use of thermal analysis techniques can realize complementarity of different characterization methods. Comprehensive thermal analysis with both thermogravimetric analysis and differential thermal analysis (TG/DTA) can measure not only mass change of a sample but also its temperature change during programmed heating-induced reaction or phase transition processes, thereby obtaining multiaspect thermal information of the material such as dehydration, structural decomposition, phase change and thermal stability. This study proposes and develops a MEMS chip-based TG/DTA microsystem that integrates both programmed heating and detecting elements into a TG chip and a DTA chip to enable the microinstrument performing TG/DTA joint characterization under microscope observation. The TG chip contains a self-heating resonant microcantilever to measure heating-induced mass change of a sample and the DTA chip is with a microheater and a temperature-detecting thermopile integrated on a suspended thermal-insulating diaphragm. Only nanogram and microgram-level samples are needed for the TG and DTA chips, thereby achieving safe measurement to energetic materials such as strong oxidants. The chip-based microinstrument surpasses the state-of-the-art commercial TG/DTA instruments that have, in the long term, suffered from large sample-amount (milligram level) requirements and have been unable to measure energetic materials. Compared with commercial instruments, the chip-based microinstrument is advantageous given its more accurate analysis, much higher heating rate, much smaller instrument volume and much lower power consumption, etc. The microinstrument has been fabricated by using wafer-level MEMS techniques. Testing results show that the mass-detection sensitivity of the TG-chip is as high as 0.45 Hz/pg in air and the temperature sensitivity of the DTA chip achieves 2.9 mV/K under the high heating rate of 25 °C/s. The strong oxidant of KMnO is analyzed with the TG/DTA joint characterization under microscopic observation. At the same time as microscope observation of the thermal decomposition phenomena, two-step thermal decomposition process of KMnO is identified and the thermal decomposition temperatures are obtained. The TG/DTA microinstrument is promising to be applied for study of various materials.
热分析技术的联合使用能够实现不同表征方法的互补。热重分析和差示热分析(TG/DTA)相结合的综合热分析不仅可以测量样品在程序升温引发的反应或相变过程中的质量变化,还能测量其温度变化,从而获取材料的多方面热信息,如脱水、结构分解、相变和热稳定性等。本研究提出并开发了一种基于MEMS芯片的TG/DTA微系统,该系统将程序升温元件和检测元件集成到TG芯片和DTA芯片中,以便在显微镜观察下实现微仪器进行TG/DTA联合表征。TG芯片包含一个自热式谐振微悬臂梁,用于测量样品在加热过程中引起的质量变化;DTA芯片则在一个悬浮的隔热膜片上集成了一个微型加热器和一个温度检测热电堆。TG和DTA芯片仅需纳克和微克级别的样品,从而实现对强氧化剂等含能材料的安全测量。基于芯片的微仪器超越了长期以来存在大样品量(毫克级别)需求且无法测量含能材料的现有商用TG/DTA仪器。与商用仪器相比,基于芯片的微仪器具有分析更准确、加热速率更高、仪器体积更小、功耗更低等优势。该微仪器采用晶圆级MEMS技术制造。测试结果表明,TG芯片在空气中的质量检测灵敏度高达0.45 Hz/pg,DTA芯片在25℃/s的高加热速率下温度灵敏度达到2.9 mV/K。在显微镜观察下,通过TG/DTA联合表征对强氧化剂KMnO₄进行了分析。在显微镜观察热分解现象的同时,确定了KMnO₄的两步热分解过程并获得了热分解温度。TG/DTA微仪器有望应用于各种材料的研究。