To assess test-retest reliability and validity of a new diagnostic device, the Shoulder Elbow Perturbator, to quantify muscle weakness, abnormal synergy, (muscle activity-related) spasticity, and changes in viscoelastic joint properties of the elbow.
Stroke patients, adults with cerebral palsy and healthy controls.
Test-retest reliability was evaluated using intra-class correlations (ICC) and assessment of measurement error. The device's validity was evaluated by demonstrating differences between patients and healthy controls, and correlations of spasticity and abnormal synergy outcomes using the clinical Modified Tardieu Scale, the Fugl-Meyer Assessment, and the Test of Arm Selective Control.
Reliability was excellent, with an ICC > 0.75 for synergy and ICCs > 0.90 for all other impairments, with relatively small measurement errors. Validity was confirmed by group differences between patients and healthy controls for muscle weakness, spasticity, and viscoelastic joint properties, but not for abnormal synergy. Correlation analysis with clinical scales confirmed validity for spasticity, while, for synergy, correlations were found in the patients with stroke, but not those with cerebral palsy.
This new diagnostic device is a reliable and valid instrument to assess multiple upper limb impairments in patients with neurological conditions, supporting its use in clinical practice. Further studies are needed to confirm these findings.
评估一种新的诊断设备——肩部肘部干扰器(Shoulder Elbow Perturbator)在量化肌肉无力、异常协同作用(肌肉活动相关)痉挛和肘部粘弹性关节特性变化方面的测试-重测信度和效度。
中风患者、脑瘫成人和健康对照者。
采用组内相关系数(ICC)和测量误差评估来评估测试-重测信度。通过证明患者与健康对照组之间的差异,以及使用临床改良 Tardieu 量表、Fugl-Meyer 评估和手臂选择性控制测试评估痉挛和异常协同作用结果之间的相关性,来评估设备的有效性。
信度非常好,协同作用的 ICC > 0.75,所有其他损伤的 ICC > 0.90,测量误差相对较小。通过患者与健康对照组之间的组间差异,确认了肌肉无力、痉挛和粘弹性关节特性的有效性,但对异常协同作用则没有。与临床量表的相关分析证实了痉挛的有效性,而对于协同作用,仅在中风患者中存在相关性,而在脑瘫患者中则没有。
这种新的诊断设备是一种可靠且有效的评估工具,可用于评估神经状况患者的多种上肢损伤,支持其在临床实践中的应用。需要进一步的研究来证实这些发现。