Nicolucci Antonio, Leonardi Luca
Center for Outcomes Research and Clinical Epidemiology (CORESEARCH Srl), Pescara (PE), Italy.
Pikdare S.p.A., Casnate Con Bernate (CO), Italy.
Med Devices (Auckl). 2022 Mar 29;15:71-77. doi: 10.2147/MDER.S352849. eCollection 2022.
To test the compatibility of PiC Insupen needles with a broad range of pens produced by different manufacturers for the injection of subcutaneously administered drugs.
The "extreme" products in PiC pen needles range were considered (G33x4 mm and G29x12 mm), to verify that the compatibility was not affected by the cannula diameter and length. Following the launch of the G34x3.5mm needle, additional tests were performed comparing G33x4 and G34x3.5 pen needles. A test medium with viscosity similar to insulin was used. Additional tests were performed with a liquid with a much higher viscosity. All the requirements of the technical standard ISO 11608-2 were satisfied, and the differences between G29, G33 and G34 were negligible. Therefore, the PiC Insupen needle G33x4mm was chosen as representative of the PiC pen needles. Dose accuracy and needle hub torque were assessed, according to the ISO 11608-2:2012 norms. For pens with variable volume, two different volumes were tested (Vlow and Vhigh) in random order, testing 60 needles with Vlow and 60 with Vhigh.
Overall, 50 different pens were tested. Dose accuracy acceptance criteria were met for all the pens tested, with the only exception of Vhigh for Berlipen Precision pen. The removal torque was adequate for all pens, with the only exception of Berlipen 301 and Berlipen 302 pens.
We documented that Pikdare needles are compatible with a large array of different pens for the injection of insulin and other drugs administered subcutaneously.
测试PiC Insupen针头与不同制造商生产的多种笔式注射器在皮下注射药物时的兼容性。
考虑PiC笔式针头系列中的“极端”产品(G33x4 mm和G29x12 mm),以验证兼容性不受套管直径和长度的影响。在推出G34x3.5mm针头后,进行了额外测试,比较G33x4和G34x3.5笔式针头。使用了一种粘度与胰岛素相似的测试介质。还用一种粘度高得多的液体进行了额外测试。所有技术标准ISO 11608-2的要求均得到满足,G29、G33和G34之间的差异可忽略不计。因此,选择PiC Insupen针头G33x4mm作为PiC笔式针头的代表。根据ISO 11608-2:2012规范评估剂量准确性和针座扭矩。对于可变剂量的笔式注射器,随机测试两个不同剂量(Vlow和Vhigh),测试60个Vlow剂量的针头和60个Vhigh剂量的针头。
总体而言,测试了50种不同的笔式注射器。除了Berlipen Precision笔式注射器的Vhigh剂量外,所有测试的笔式注射器均符合剂量准确性验收标准。除了Berlipen 301和Berlipen 302笔式注射器外,所有笔式注射器的拔出扭矩都足够。
我们证明Pikdare针头与大量不同的笔式注射器兼容,可用于注射胰岛素和其他皮下给药的药物。