Abstract

BACKGROUND

The pen injection format, typically used for insulin administration, has been adapted for the injectable, noninsulin diabetes therapy pramlintide. Administered before major meals, pramlintide therapy requires two to four injections/day in addition to the patients' usual insulin injections. The dose accuracy and injection force was determined for the 60 and 120 µg pramlintide pens.

METHODS

Dose accuracy testing was conducted at two sites on multiple 60 µg (15, 30, and 60 µg doses) and 120 µg pens (60 and 120 µg doses) at prespecified temperatures (5-40 °C) and humidities (0-75%) using 29 G half-inch needles. All pens were stabilized under testing conditions for 4 h prior to testing. One site used a compression load cell (Zwick device) to test pens; one site performed tests manually. Injection-force testing was conducted at one site on multiple 60 and 120 µg pens at multiple temperatures (18-28 °C) and humidities (25-75%) using 29 and 31 G half-inch needles and an injection speed of 150 m/min. Injection-force testing was performed using a Zwick device.

RESULTS

Dose accuracy for all pens tested, regardless of location, reproducibly met/exceeded acceptance criteria. Mean percentage of dose accuracy was 96.04 to 102.45% [standard deviations (SDs) 0.3 to 1.4 µg] for the 60 µg pen and 98.16 to 101.83% (SDs 0.4 to 2.5 µg) for the 120 µg pen. The average injection force across both pens did not exceed 7 N regardless of needle size.

CONCLUSIONS

High dose accuracy and low injection force were observed for the 60 and 120 µg pens under a variety of conditions.

摘要

背景

笔式注射形式通常用于胰岛素给药,现已应用于可注射的非胰岛素糖尿病治疗药物普兰林肽。普兰林肽治疗需在主餐前给药,除患者常规胰岛素注射外,每天还需注射两到四次。本研究测定了60微克和120微克普兰林肽笔的剂量准确性和注射力。

方法

在两个地点,使用29G半英寸针头,于预设温度(5 - 40°C)和湿度(0 - 75%)条件下,对多支60微克(15微克、30微克和60微克剂量)和120微克笔(60微克和120微克剂量)进行剂量准确性测试。所有笔在测试前于测试条件下稳定4小时。一个地点使用压缩力传感器(Zwick装置)测试笔;另一个地点手动进行测试。在一个地点,使用29G和31G半英寸针头,注射速度为150米/分钟,于多个温度(18 - 28°C)和湿度(25 - 75%)条件下,对多支60微克和120微克笔进行注射力测试。注射力测试使用Zwick装置进行。

结果

无论在哪个地点,所有测试笔的剂量准确性均可重复地达到/超过验收标准。60微克笔的剂量准确性平均百分比为96.04%至·102.45%[标准差(SDs)为0.3至1.4微克],120微克笔为98.16%至101.83%(SDs为0.4至2.5微克)。无论针头尺寸如何,两种笔的平均注射力均未超过7牛。

结论

在各种条件下,60微克和120微克笔均表现出高剂量准确性和低注射力。

相似文献

5
Insulin pens.胰岛素笔
Diabetes Forecast. 2014 Jan;67(1):68-9.
9
10
Study on the dosing accuracy of commonly used disposable insulin pens.常用一次性胰岛素笔注射剂量准确性的研究。
Diabetes Technol Ther. 2012 Sep;14(9):804-9. doi: 10.1089/dia.2011.0298. Epub 2012 Jun 13.

本文引用的文献

2
Practical aspects of insulin pen devices.胰岛素笔装置的实际应用方面。
J Diabetes Sci Technol. 2010 May 1;4(3):522-31. doi: 10.1177/193229681000400304.
3
Evolution of diabetes insulin delivery devices.糖尿病胰岛素输送装置的演变
J Diabetes Sci Technol. 2010 May 1;4(3):505-13. doi: 10.1177/193229681000400302.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验