Kwok Wingchi E
From the Department of Imaging Sciences, University of Rochester, 601 Elmwood Ave, Rochester, NY 14642; and University of Rochester Center for Advanced Brain Imaging and Neurophysiology, Rochester, NY.
Radiographics. 2022 May-Jun;42(3):898-918. doi: 10.1148/rg.210110. Epub 2022 Apr 8.
Radiofrequency (RF) coils are an essential MRI component used for transmission of the RF field to excite nuclear spins and for reception of the MRI signal. They play an important role in image quality in terms of signal-to-noise ratio, signal uniformity, and image resolution. However, they are also associated with potential image artifacts and RF heating that may lead to patient burns. Knowledge of the basic principles of RF coils-including coil designs commonly used in clinical MRI and the anatomy of RF receive coils-facilitates understanding of the use and safety issues of RF coils. Selection of suitable RF coils for individual applications and proper use of RF coils in particular MRI techniques such as parallel imaging are needed to achieve optimal image quality, prevent image artifacts, and reduce the risk of RF burns. The ability to correctly identify RF coil problems and distinguish them from other problems with image artifacts resembling those of RF coil problems allows effective handling of the problems and efficient clinical MRI operation. Quality control of RF coils is required to ensure consistent image quality for clinical MRI and avoid coil problems that may affect image diagnostic evaluation or interrupt patient imaging. There are different phantom test methods for RF coil quality control; the appropriate one to use depends on the coil design and MRI system. RSNA, 2022.
射频(RF)线圈是磁共振成像(MRI)的重要组成部分,用于传输射频场以激发核自旋并接收MRI信号。它们在图像质量方面,包括信噪比、信号均匀性和图像分辨率,发挥着重要作用。然而,它们也与潜在的图像伪影和射频加热相关,这可能导致患者灼伤。了解RF线圈的基本原理,包括临床MRI中常用的线圈设计以及RF接收线圈的解剖结构,有助于理解RF线圈的使用和安全问题。为了获得最佳图像质量、防止图像伪影并降低射频灼伤风险,需要为个别应用选择合适的RF线圈,并在特定的MRI技术(如并行成像)中正确使用RF线圈。正确识别RF线圈问题并将其与其他类似RF线圈问题的图像伪影问题区分开来的能力,有助于有效处理这些问题并实现高效的临床MRI操作。需要对RF线圈进行质量控制,以确保临床MRI的图像质量一致,并避免可能影响图像诊断评估或中断患者成像的线圈问题。对于RF线圈质量控制有不同的体模测试方法;应使用的合适方法取决于线圈设计和MRI系统。RSNA,2022年。